Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons
Medical Physics · 2024
DOI: 10.1002/mp.16925 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Ashley J Cetnar und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Electrons, Particle Accelerators. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Varian Medical Systems.
Abstract (Originalsprache)
Background
Ultra‐high dose rate radiation (UHDR) is being explored by researchers in promise of advancing radiation therapy treatments.
Purpose
This work presents the commissioning of Varian's Flash Extension for research (FLEX) conversion of a Clinac to deliver UHDR electrons.
Methods
A Varian Clinac iX with the FLEX conversion was commissioned for non‐clinical research use with 16 MeV UHDR (16H) energy. This involved addition of new hardware, optimizing the electron gun voltages, radiofrequency (RF) power, and steering coils in order to maximize the accelerated electron beam current, sending the beam through custom scattering foils to produce the UHDR with 16H beam. Profiles and percent depth dose (PDD) measurements for 16H were obtained using radiochromic film in a custom vertical film holder and were compared to 16 MeV conventional electrons (16C). Dose rate and dose per pulse (DPP) were calculated from measured dose in film. Linearity and stability were assessed using an Advanced Markus ionization chamber.
Results
Energies for 16H and 16C had similar beam quality based on PDD measurements. Measurements at the head of the machine (61.3 cm SSD) with jaws set to 10×10 cm 2 showed the FWHM of the profile as 7.2 cm, with 3.4 Gy as the maximum DPP and instantaneous dose rate of 8.1E5 Gy/s. Measurements at 100 cm SSD with 10 cm standard cone showed the full width at half max (FWHM) of the profile as 10.5 cm, 1.08 Gy as the maximum DPP and instantaneous dose rate of 2.E5 Gy/s. Machine output with number of pulses was linear ( R = 1) from 1 to 99 delivered pulses. Output stability was measured within ±1% within the same session and within ±2% for daily variations.
Conclusions
The FLEX conversion of the Clinac is able to generate UHDR electron beams which are reproducible with beam properties similar to clinically used electrons at 16 MeV. Having a platform which can quickly transition between UHDR and conventional modes (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, James Cancer Hospital and Solove Research Institute, The Ohio State University, Columbus, Ohio, USA.
Förderung
- Varian Medical Systems
Zitieren
APA
Cetnar, A. J., Jain, S., Gupta, N., & Chakravarti, A. (2024). Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons. Medical Physics. https://doi.org/10.1002/mp.16925
Vancouver
Cetnar AJ, Jain S, Gupta N, Chakravarti A. Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons. Medical Physics. 2024. doi:10.1002/mp.16925
BibTeX
@article{Cetnar2024,
author = {Cetnar, Ashley J and Jain, Sagarika and Gupta, Nilendu and Chakravarti, Arnab},
title = {Technical note: Commissioning of a linear accelerator producing ultra‐high dose rate electrons},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.16925}
}
EUVIMED