Journal article in Medical Physics, published online on 2024-01-03. Written by Laurens Beljaards and 5 co-authors. Institution: Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. Major MEDLINE topics: Artifacts, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Movement. Cited 9 times so far (Crossref); 25 references. The full text is freely available (open access, CC BY 4.0). Funded by: Dutch Research Council (NWO), Philips Research, ROBUST: Trustworthy AI-Based Systems for Sustainable Growth.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Magnetic Resonance acquisition is a time consuming process, making it susceptible to patient motion during scanning. Even motion in the order of a millimeter can introduce severe blurring and ghosting artifacts, potentially necessitating re‐acquisition. Magnetic Resonance Imaging (MRI) can be accelerated by acquiring only a fraction of k‐space, combined with advanced reconstruction techniques leveraging coil sensitivity profiles and prior knowledge. Artificial intelligence (AI)‐based reconstruction techniques have recently been popularized, but generally assume an ideal setting without intra‐scan motion.

Purpose

To retrospectively detect and quantify the severity of motion artifacts in undersampled MRI data. This may prove valuable as a safety mechanism for AI‐based approaches, provide useful information to the reconstruction method, or prompt for re‐acquisition while the patient is still in the scanner.

Methods

We developed a deep learning approach that detects and quantifies motion artifacts in undersampled brain MRI. We demonstrate that synthetically motion‐corrupted data can be leveraged to train the convolutional neural network (CNN)‐based motion artifact estimator, generalizing well to real‐world data. Additionally, we leverage the motion artifact estimator by using it as a selector for a motion‐robust reconstruction model in case a considerable amount of motion was detected, and a high data consistency model otherwise.

Results

Training and validation were performed on 4387 and 1304 synthetically motion‐corrupted images and their uncorrupted counterparts, respectively. Testing was performed on undersampled in vivo motion‐corrupted data from 28 volunteers, where our model distinguished head motion from motion‐free scans with 91% and 96% accuracy when trained on synthetic and on real data, respectively. It predicted a manually defined quality label (‘Good’, ‘Medium’ or ‘Bad’ quality) correctly in 76% and 85% of the time when trained on synthetic and real data, respectively. When used as a selector it selected the appropriate reconstruction network 93% of the time, achieving near optimal SSIM values.

Conclusions

The proposed method quantified motion artifact severity in undersampled MRI data with high accuracy, enabling real‐time motion artifact detection that can help improve the safety and quality of AI‐based reconstructions.

Subjects

MeSH terms (MEDLINE)

ArtifactsMagnetic Resonance ImagingImage Processing, Computer-AssistedMovementHumansArtificial IntelligenceBrainDeep Learning

Author keywords

MRI reconstructionaccelerated MRIdeep learningmotion artifact severity estimationmotion corruption

Authors and institutions

  • Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.
  • Cardiologs, Philips, Paris, France.
  • Faculty of Computer Science, Eindhoven University of Technology, Eindhoven, The Netherlands.
  • Philips Research, Hamburg, Germany.

Funding

  • Dutch Research Council (NWO)
  • Philips Research
  • ROBUST: Trustworthy AI-Based Systems for Sustainable Growth
  • Dutch Ministry of Economic Affairs and Climate Policy (EZK)

Cite

APA

Beljaards, L., Pezzotti, N., Rao, C., Doneva, M., van Osch, M. J. P., & Staring, M. (2024). AI‐based motion artifact severity estimation in undersampled MRI allowing for selection of appropriate reconstruction models. Medical Physics, 51(5), 3555-3565. https://doi.org/10.1002/mp.16918

Vancouver

Beljaards L, Pezzotti N, Rao C, Doneva M, van Osch MJP, Staring M. AI‐based motion artifact severity estimation in undersampled MRI allowing for selection of appropriate reconstruction models. Medical Physics. 2024;51(5):3555-3565. doi:10.1002/mp.16918

BibTeX
@article{Beljaards2024,
  author = {Beljaards, Laurens and Pezzotti, Nicola and Rao, Chinmay and Doneva, Mariya and van Osch, Matthias J P and Staring, Marius},
  title = {AI‐based motion artifact severity estimation in undersampled MRI allowing for selection of appropriate reconstruction models},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {5},
  pages = {3555--3565},
  doi = {10.1002/mp.16918}
}
Start for free →