EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Qualitative and quantitative analysis of functional cardiac MRI using a novel compressed SENSE sequence with artificial intelligence image reconstruction

Konstantin Klein, Marcel Christian Langenbach, Lenhard Pennig, Thomas Schömig, Robert Terzis, Isabel Luisa Langenbach, David Maintz, Matej Gajzler, Kristina Sonnabend, Claas Philip Nähle

Magnetic Resonance Imaging · 2025

DOI: 10.1016/j.mri.2025.110448 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2025. Verfasst von Konstantin Klein und 9 weiteren Autor:innen. Beteiligte Einrichtung: University Hospital Cologne, Cologne, Germany. Electronic address: konstantin.klein1@uk-koeln.de. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Heart, Magnetic Resonance Imaging, Cine, Artificial Intelligence. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: To evaluate the feasibility of combining Compressed SENSE (CS) with a newly developed deep learning-based algorithm (CS-AI) using a Convolutional Neural Network to accelerate balanced steady-state free precession (bSSFP)-sequences for cardiac magnetic resonance imaging (MRI). METHODS: 30 healthy volunteers were examined prospectively with a 3 T MRI scanner. We acquired CINE bSSFP sequences for short axis (SA, multi-breath-hold) and four-chamber (4CH)-view of the heart. For each sequence, four different CS accelerations and CS-AI reconstructions with three different denoising parameters, CS-AI medium, CS-AI strong, and CS-AI complete, were used. Cardiac left ventricular (LV) function (i.e., ejection fraction, end-diastolic volume, end-systolic volume, and LV mass) was analyzed using the SA sequences in every CS factor and each AI level. Two readers, blinded to the acceleration and denoising levels, evaluated all sequences regarding image quality and artifacts using a 5-point Likert scale. Friedman and Dunn's multiple comparison tests were used for qualitative evaluation, ANOVA and Tukey Kramer test for quantitative metrics. RESULTS: Scan time could be decreased up to 57 % for the SA-Sequences and up to 56 % for the 4CH-Sequences compared to the clinically established sequences consisting of SA-CS3 and 4CH-CS2,5 (SA-CS3: 112 s vs. SA-CS6: 48 s; 4CH-CS2,5: 9 s vs. 4CH-CS5: 4 s, p 0.05). The image quality loss and artifact increase accompanying increasing acceleration levels could be entirely compensated by CS-AI post-processing, with the best results for image quality using the combination of the highest CS factor with strong AI (SA-CINE: Coef.:1.31, 95 %CI:1.05-1.58; 4CH-CINE: Coef.:1.18, 95 %CI:1.05-1.58; both p < 0.001), and with complete AI regarding the artifact score (SA-CINE: Coef.:1.33, 95 %CI:1.06-1.60; 4CH-CINE: Coef.:1.31, 95 %CI:0.86-1.77; both p < 0.001). CONCLUSION: Combining CS sequences with AI-based image reconstruction for denoising significantly decreases scan time in cardiac imaging while upholding LV functional analysis accuracy and delivering stable outcomes for image quality and artifact reduction. This integration presents a promising advancement in cardiac MRI, promising improved efficiency without compromising diagnostic quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Processing, Computer-AssistedHeartMagnetic Resonance Imaging, CineArtificial IntelligenceMagnetic Resonance ImagingHumansMaleFemaleAdultAlgorithmsProspective StudiesHealthy VolunteersArtifactsMiddle AgedYoung AdultDeep LearningImage Interpretation, Computer-AssistedFeasibility StudiesVentricular Function, LeftNeural Networks, ComputerReproducibility of Results

Stichwörter der Autor:innen

Accelerated MRI scansArtificial intelligenceCardiac MRIDeep learningMagnetic resonance imaging

Autor:innen und Institutionen

  • University Hospital Cologne, Cologne, Germany. Electronic address: konstantin.klein1@uk-koeln.de.
  • University Hospital Cologne, Cologne, Germany; Cardiovascular Imaging Research Center, Massachusetts General Hospital, Boston, MA, USA. Electronic address: mlangenbach@mgh.harvard.edu.
  • University Hospital Cologne, Cologne, Germany. Electronic address: lenhard.pennig@uk-koeln.de.
  • University Hospital Cologne, Cologne, Germany. Electronic address: thomas.schoemig@uk-koeln.de.
  • University Hospital Cologne, Cologne, Germany. Electronic address: robert.terzis@uk-koeln.de.
  • University Hospital Cologne, Cologne, Germany; Cardiovascular Imaging Research Center, Massachusetts General Hospital, Boston, MA, USA. Electronic address: ILangenbach@mgh.harvard.edu.

Zitieren

APA

Klein, K., Langenbach, M. C., Pennig, L., Schömig, T., Terzis, R., Langenbach, I. L., Maintz, D., Gajzler, M., Sonnabend, K., & Nähle, C. P. (2025). Qualitative and quantitative analysis of functional cardiac MRI using a novel compressed SENSE sequence with artificial intelligence image reconstruction. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110448

Vancouver

Klein K, Langenbach MC, Pennig L, Schömig T, Terzis R, Langenbach IL, et al. Qualitative and quantitative analysis of functional cardiac MRI using a novel compressed SENSE sequence with artificial intelligence image reconstruction. Magnetic Resonance Imaging. 2025. doi:10.1016/j.mri.2025.110448

BibTeX
@article{Klein2025,
  author = {Klein, Konstantin and Langenbach, Marcel Christian and Pennig, Lenhard and Schömig, Thomas and Terzis, Robert and Langenbach, Isabel Luisa and Maintz, David and Gajzler, Matej and Sonnabend, Kristina and Nähle, Claas Philip},
  title = {Qualitative and quantitative analysis of functional cardiac MRI using a novel compressed SENSE sequence with artificial intelligence image reconstruction},
  journal = {Magnetic Resonance Imaging},
  year = {2025},
  doi = {10.1016/j.mri.2025.110448}
}
Kostenlos starten →