Signal‐difference‐to‐noise comparison of temporal subtraction, kV‐switching dual‐energy and photon‐counting dual‐energy x‐ray angiography

Sarah Aubert, Jesse Tanguay

Medical Physics · 2023-12 · 50 (12) · 7400-7414

DOI: 10.1002/mp.16800 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 25.10.2023. Verfasst von Sarah Aubert und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Toronto Metropolitan University (formerly Ryerson University), Toronto, Canada. Hauptthemen laut MEDLINE: Angiography, Iodine. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada, Canadian Foundation for Innovation, Ontario Research Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Dual‐energy (DE) x‐ray angiography with photon‐counting detectors (PCDs) may enable single‐exposure DE imaging of coronary vasculature.

Purpose

To compare the iodine signal‐difference‐to‐noise ratio (SDNR) of single‐exposure DE angiography with digital subtraction angiography (DSA) and kV‐switching DE angiography for matched patient x‐ray exposure.

Methods

In a phantom study, we determined the technique parameters that maximized the iodine SDNR per root entrance air kerma for DSA, kV‐switching DE angiography and single‐exposure DE angiography. We measured SDNR from images of a phantom consisting of an iodine step‐wedge immersed in a water tank of either 20 or 30 cm in thickness. We also imaged a phantom with simulated vessels embedded in background clutter and measured vessel SDNR. For this second phantom, we also applied anti‐correlated noise reduction (ACNR) and calculated the resulting iodine SDNR. All images were acquired using a cadmium telluride PCD with two energy bins and analog charge summing for charge sharing suppression. The energy‐discrimination capabilities were only used for the single‐exposure DE approach. Optimized techniques were compared in terms of SDNR per root air kerma for two levels of x‐ray scatter.

Results

For the same patient x‐ray exposure, the SDNR of single‐exposure DE imaging without ACNR was 75% to 85% of that of kV‐switching DE imaging (also without ACNR) and DSA, the latter two of which had nearly equal SDNR. The single‐exposure DE approach required ∼50% of the tube load of the kV‐switching approach to achieve the same SDNR. For matched patient air kermas, the single exposure approach required only ∼25% of the tube load of the kV‐switching approach. ACNR increased SDNR by 2.4 and 3.0 for kV‐switching and single‐exposure DE imaging, respectively.

Conclusions

Photon‐counting, single‐exposure DE angiography can suppress soft tissues and provide iodine SDNR levels comparable to DSA and kV‐switching DE angiography for matched patient radiation exposures. When ACNR is used to reduce DE image noise, the SDNR of single‐exposure DE imaging and kV‐switching DE imaging exceed that of DSA by more than a factor of two. Compared to kV‐switching DE imaging, single‐exposure DE imaging requires substantially lower tube loading to achieve the same SDNR.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AngiographyIodineHumansX-RaysRadiographySignal-To-Noise RatioPhantoms, Imaging

Stichwörter der Autor:innen

angiographydual-energy imagingimage qualityphoton countingspectral imagingvascular imagingx-ray imaging

Substanzen

Iodine

Autor:innen und Institutionen

  • Department of Physics, Toronto Metropolitan University (formerly Ryerson University), Toronto, Canada.

Förderung

  • Natural Sciences and Engineering Research Council of Canada
  • Canadian Foundation for Innovation
  • Ontario Research Foundation

Zitieren

APA

Aubert, S., & Tanguay, J. (2023). Signal‐difference‐to‐noise comparison of temporal subtraction, kV‐switching dual‐energy and photon‐counting dual‐energy x‐ray angiography. Medical Physics, 50(12), 7400-7414. https://doi.org/10.1002/mp.16800

Vancouver

Aubert S, Tanguay J. Signal‐difference‐to‐noise comparison of temporal subtraction, kV‐switching dual‐energy and photon‐counting dual‐energy x‐ray angiography. Medical Physics. 2023;50(12):7400-7414. doi:10.1002/mp.16800

BibTeX
@article{Aubert2023,
  author = {Aubert, Sarah and Tanguay, Jesse},
  title = {Signal‐difference‐to‐noise comparison of temporal subtraction, kV‐switching dual‐energy and photon‐counting dual‐energy x‐ray angiography},
  journal = {Medical Physics},
  year = {2023},
  volume = {50},
  number = {12},
  pages = {7400--7414},
  doi = {10.1002/mp.16800}
}
Kostenlos starten →