Comparison of Low‐Iodine Concentration Quantification Accuracy and Contrast‐To‐Noise Ratio Across Three Dual‐Energy CT Techniques: A Phantom Study

Chonsitron Sirituenlee, Sirapasson Kraison, Paweena Wongpan, Pawitiya Supawarapong, Teeraphat Pataraphitakpong, Supannika Kawvised, Samrit Kittipayak, Monchai Phonlakrai

Journal of Medical Radiation Sciences · 2026

DOI: 10.1002/jmrs.70070 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Medical Radiation Sciences, 2026. Verfasst von Chonsitron Sirituenlee und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Radiological Technology, Faculty of Health Science Technology, Chulabhorn Royal Academy, Bangkok, Thailand. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Iodine, Contrast Media. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Dual‐energy computed tomography (DECT) enables improved tissue characterisation and the potential to reduce contrast media dose, yet clinical practice often still uses contrast volumes comparable to single‐energy CT. This study evaluates the accuracy of low‐concentration iodine quantification and image quality across three DECT techniques, with specific objectives to identify the lowest reliably measurable iodine concentration and to compare normalised contrast‐to‐noise ratios (CNR norm ) using a clinical abdominal CT protocol.

Methods

A Multi‐Energy CT Quality Assurance Phantom (CIRS Model 662) filled with iodinated contrast (Iopamiro 370 mgI/mL) at concentrations of 0.025–10 mgI/mL was scanned three times using the clinical abdominal imaging protocol at the study site, without parameter adjustment. Iodine maps and virtual monoenergetic images (VMI) were analyzed using regions of interest to evaluate iodine measurement accuracy for limit of detection (LOD) and limit of quantification (LOQ) and normalised CNR across the DECT techniques.

Results

For VMI at 40 keV, all DECT techniques demonstrated high iodine measurement accuracy ( R 2 > 0.99). The LODs for rapid kV‐switching, dual‐layer detector, and dual‐source DECT were 0.205, 0.436 and 0.411 mg/mL, respectively. Corresponding LOQs were 0.621, 1.321 and 1.246 mg/mL. The rapid kV‐switching system consistently achieved the lowest LOD and LOQ, while the dual‐layer detector yielded the highest CNR norm , followed by dual‐source DECT and rapid kV‐switching.

Conclusions

All DECT techniques accurately quantified iodine on iodine maps. Rapid kV‐switching provided the lowest LOD and LOQ, whereas the dual‐layer detector offered superior image quality in terms of CNR norm derived from VMI at 40 keV in CT abdominal protocol setting.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedIodineContrast MediaSignal-To-Noise RatioHumans

Stichwörter der Autor:innen

CT performancecomputed tomographycontrast mediadual‐energy CTiodine concentrationiodine mapvirtual monoenergetic image

Substanzen

IodineContrast Media

Autor:innen und Institutionen

  • School of Radiological Technology, Faculty of Health Science Technology, Chulabhorn Royal Academy, Bangkok, Thailand.
  • Diagnostic and Interventional Radiology Department, Chulabhorn Hospital, Chulabhorn Royal Academy, Bangkok, Thailand.

Zitieren

APA

Sirituenlee, C., Kraison, S., Wongpan, P., Supawarapong, P., Pataraphitakpong, T., Kawvised, S., Kittipayak, S., & Phonlakrai, M. (2026). Comparison of Low‐Iodine Concentration Quantification Accuracy and Contrast‐To‐Noise Ratio Across Three Dual‐Energy CT Techniques: A Phantom Study. Journal of Medical Radiation Sciences. https://doi.org/10.1002/jmrs.70070

Vancouver

Sirituenlee C, Kraison S, Wongpan P, Supawarapong P, Pataraphitakpong T, Kawvised S, et al. Comparison of Low‐Iodine Concentration Quantification Accuracy and Contrast‐To‐Noise Ratio Across Three Dual‐Energy CT Techniques: A Phantom Study. Journal of Medical Radiation Sciences. 2026. doi:10.1002/jmrs.70070

BibTeX
@article{Sirituenlee2026,
  author = {Sirituenlee, Chonsitron and Kraison, Sirapasson and Wongpan, Paweena and Supawarapong, Pawitiya and Pataraphitakpong, Teeraphat and Kawvised, Supannika and Kittipayak, Samrit and Phonlakrai, Monchai},
  title = {Comparison of Low‐Iodine Concentration Quantification Accuracy and Contrast‐To‐Noise Ratio Across Three Dual‐Energy CT Techniques: A Phantom Study},
  journal = {Journal of Medical Radiation Sciences},
  year = {2026},
  doi = {10.1002/jmrs.70070}
}
Kostenlos starten →