Spectral performance evaluation of a second‐generation spectral detector CT

Leening P Liu, Nadav Shapira, Sandra S Halliburton, Sebastian Meyer, Amy Perkins, Harold I Litt, Hans Ulrich Kauczor, Tim Leiner, Wolfram Stiller, Peter B Noël

Journal of Applied Clinical Medical Physics · 2024-04 · 25 (4)

DOI: 10.1002/acm2.14300 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 22.02.2024. Verfasst von Leening P Liu und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Iodine. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to characterize a second‐generation wide‐detector dual‐layer spectral computed tomography (CT) system for material quantification accuracy, acquisition parameter and patient size dependencies, and tissue characterization capabilities.

Methods

A phantom with multiple tissue‐mimicking and material‐specific inserts was scanned with a dual‐layer spectral detector CT using different tube voltages, collimation widths, radiation dose levels, and size configurations. Accuracy of iodine density maps and virtual monoenergetic images (MonoE) were investigated. Additionally, differences between conventional and MonoE 70 keV images were calculated to evaluate acquisition parameter and patient size dependencies. To demonstrate material quantification and differentiation, liver‐mimicking inserts with adipose and iron were analyzed with a two‐base decomposition utilizing MonoE 50 and 150 keV, and root mean square error (RMSE) for adipose and iron content was reported.

Results

Measured inserts exhibited quantitative accuracy across a wide range of MonoE levels. MonoE 70 keV images demonstrated reduced dependence compared to conventional images for phantom size (1 vs. 27 HU) and acquisition parameters, particularly tube voltage (4 vs. 37 HU). Iodine density quantification was successful with errors ranging from ‐0.58 to 0.44 mg/mL. Similarly, inserts with different amounts of adipose and iron were differentiated, and the small deviation in values within inserts corresponded to a RMSE of 3.49 ± 1.76% and 1.67 ± 0.84 mg/mL for adipose and iron content, respectively.

Conclusion

The second‐generation dual‐layer CT enables acquisition of quantitatively accurate spectral data without compromises from differences in patient size and acquisition parameters.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedIodineHumansSignal-To-Noise RatioPhantoms, ImagingObesityIron

Stichwörter der Autor:innen

dual energy CTspectral CT

Substanzen

IodineIron

Autor:innen und Institutionen

  • Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Philips Healthcare, Orange Village, Ohio, USA.
  • Diagnostic and Interventional Radiology (DIR), Heidelberg University Hospital, Heidelberg, Germany.
  • Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Liu, L. P., Shapira, N., Halliburton, S. S., Meyer, S., Perkins, A., Litt, H. I., Kauczor, H. U., Leiner, T., Stiller, W., & Noël, P. B. (2024). Spectral performance evaluation of a second‐generation spectral detector CT. Journal of Applied Clinical Medical Physics, 25(4). https://doi.org/10.1002/acm2.14300

Vancouver

Liu LP, Shapira N, Halliburton SS, Meyer S, Perkins A, Litt HI, et al. Spectral performance evaluation of a second‐generation spectral detector CT. Journal of Applied Clinical Medical Physics. 2024;25(4). doi:10.1002/acm2.14300

BibTeX
@article{Liu2024,
  author = {Liu, Leening P and Shapira, Nadav and Halliburton, Sandra S and Meyer, Sebastian and Perkins, Amy and Litt, Harold I and Kauczor, Hans Ulrich and Leiner, Tim and Stiller, Wolfram and Noël, Peter B},
  title = {Spectral performance evaluation of a second‐generation spectral detector CT},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {4},
  doi = {10.1002/acm2.14300}
}
Kostenlos starten →