Model observer performance in contrast‐enhanced lesions in breast CT: The influence of contrast concentration on detectability

Su Hyun Lyu, Andrew M Hernandez, Shadi Aminololama Shakeri, Craig K Abbey, John M Boone

Medical Physics · 2023

DOI: 10.1002/mp.16667 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Su Hyun Lyu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of California Davis, Davis, California, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Iodine. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIBIB NIH HHS, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The use of iodine‐based contrast agent for better delineation of tumors in breast CT (bCT) has been shown to be compelling, similar to the tumor enhancement in contrast‐enhanced breast MRI. Contrast‐enhanced bCT (CE‐bCT) is a relatively new tool, and a structured evaluation of different imaging parameters at play has yet to be conducted. In this investigation, data sets of acquired bCT images from 253 patients imaged at our institution were used in concert with simulated mathematically inserted spherical contrast‐enhanced lesions to study the role of contrast enhancement on detectability.

Purpose

To quantitatively evaluate the improvement in lesion detectability due to contrast enhancement across lesion diameter, section thickness, view plane, and breast density using a pre‐whitened matched filter (PWMF) model observer.

Methods

The relationship between iodine concentration and Hounsfield units (HU) was measured using spectral modeling. The lesion enhancement from clinical CE‐bCT images in 22 patients was evaluated, and the average contrast enhancement (ΔHU) was determined. Mathematically generated spherical mass lesions of varying diameters (1, 3, 5, 9, 11, 15 mm) and contrast enhancement levels (0, 0.25, 0.50, 0.75, 1) were inserted at random locations in 253 actual patient bCT datasets. Images with varying thicknesses (0.4–19.8 mm) were generated by slice averaging, and the role of view plane (coronal and axial planes) was studied. A PWMF was used to generate receiver operating characteristic (ROC) curves across parameters of lesion diameter, contrast enhancement, section thickness, view plane, and breast density. The area under the ROC curve (AUC) was used as the primary performance metric, generated from over 90,000 simulated lesions.

Results

An average 20% improvement (ΔAUC = 0.1) in lesion detectability due to contrast enhancement was observed across lesion diameter, section thickness, breast density, and view plane. A larger improvement was observed when stratifying patients based on breast density. For patients with VGF ≤ 40%, detection performance improved up to 20% (until AUC →1), and for patients with denser breasts (VGF > 40%), detection performance improved more drastically, ranging from 20% to 80% for 1‐ and 5‐mm lesions. For the 1 mm lesion, detection performance raised slightly at the 1.2 mm section thickness before falling off as thickness increased. For larger lesions, detection performance was generally unaffected as section thickness increased up until it reached 5.8 mm, where performance began to decline. Detection performance was higher in the axial plane compared to the coronal plane for smaller lesions and thicker sections.

Conclusions

For emerging diagnostic tools like CE‐bCT, it is important to optimize imaging protocols for lesion detection. In this study, we found that intravenous contrast can be used to detect small lesions in dense breasts. Optimal section thickness for detectability has dependencies on breast density and lesion size, therefore, display thickness should be adjusted in real‐time using display software. These findings may be useful for the development of CE‐bCT as well as other x‐ray‐based breast imaging modalities.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedIodineHumansBreastImaging, Three-DimensionalMammographyPhantoms, Imaging

Stichwörter der Autor:innen

breast imagingcone beam CTmodel observers

Substanzen

Iodine

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of California Davis, Davis, California, USA.
  • Department of Radiology, University of California Davis, Sacramento, California, USA.
  • Department of Psychological and Brain Sciences, UC Santa Barbara, Santa Barbara, California, USA.

Förderung

  • NCI NIH HHS
  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Lyu, S. H., Hernandez, A. M., Shakeri, S. A., Abbey, C. K., & Boone, J. M. (2023). Model observer performance in contrast‐enhanced lesions in breast CT: The influence of contrast concentration on detectability. Medical Physics. https://doi.org/10.1002/mp.16667

Vancouver

Lyu SH, Hernandez AM, Shakeri SA, Abbey CK, Boone JM. Model observer performance in contrast‐enhanced lesions in breast CT: The influence of contrast concentration on detectability. Medical Physics. 2023. doi:10.1002/mp.16667

BibTeX
@article{Lyu2023,
  author = {Lyu, Su Hyun and Hernandez, Andrew M and Shakeri, Shadi Aminololama and Abbey, Craig K and Boone, John M},
  title = {Model observer performance in contrast‐enhanced lesions in breast CT: The influence of contrast concentration on detectability},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16667}
}
Kostenlos starten →