Design and imaging performance of BPET–DBT – a dedicated scanner for breast imaging
Medical Physics · 2026
DOI: 10.1002/mp.70651 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2026. Verfasst von Srilalan Krishnamoorthy und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA. Hauptthemen laut MEDLINE: Positron-Emission Tomography, Breast, Mammography. Die Arbeit nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, Burroughs Wellcome Foundation, NIH HHS.
Abstract (Originalsprache)
Background
A dedicated breast PET (BPET) scanner integrated with digital breast tomosynthesis (DBT) enables three‐dimensional functional imaging with quantitative information co‐registered to high‐resolution anatomical images. With the availability of breast cancer specific PET radiotracers, such a scanner has the potential to improve treatment planning in breast cancer. We recently developed a BPET–DBT scanner utilizing an integrated gantry design to generate intrinsically co‐registered BPET–DBT images.
Purpose
This paper presents the scanner design, characterizes its imaging performance, and reports on our initial human imaging experience to demonstrate scanner's clinical readiness.
Methods
The PET component is comprised of two detector heads, providing a 20 × 10 × 10 cm 3 scanner field‐of‐view. The detector provides time‐of‐flight (TOF) measurement and is comprised of 32×32 LYSO arrays of 1.5 × 1.5 × 15 mm 3 crystals coupled to multi‐anode PMTs. Custom waveform‐sampling electronics ensures high spatial and timing resolution with minimal deadtime. The DBT component is a state‐of‐the‐art system which employs 2D x‐ray tube and detector motion for improved contrast and resolution. Phantoms appropriate for breast imaging were imaged to assess spatial resolution, image quantitation, count‐rate capability, and co‐registration accuracy. First human imaging was performed in volunteers with estrogen‐receptor‐positive (ER+) breast cancer. A five‐minute BPET–DBT scan using 18 F‐fluoroestradiol was acquired and compared with a whole‐body PET scan.
Results
System timing resolution of 425 ps and energy resolution of 16% were measured. Spatial resolution measurements showed an in‐plane (parallel to the detector) resolution of 2 mm and demonstrates the ability to resolve 1.6 mm rods in the micro‐deluxe hot rod phantom. TOF reconstruction combined with image‐based resolution modeling mitigates artifacts typical in scanners with incomplete angular coverage and enables good out‐of‐plane (orthogonal to detector) discrimination of the 2.4 mm rods. Count‐rate measurements confirm low scanner dead‐time, with sufficient count‐rate capability for clinical use.
Conclusions
We successfully constructed and tested a BPET scanner and integrated it with DBT. BPET scanner performance was characterized, and imaging capabilities relevant for clinical imaging have been highlighted. Images from a patient with ER+ breast cancer show clear tumor delineation and visualization of heterogeneous uptake, as also seen with WB‐PET. This approach offers a promising strategy for improved treatment planning and outcomes in breast cancer care.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Physics and Astronomy, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Abramson Cancer Center, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Förderung
- NIBIB NIH HHS
- Burroughs Wellcome Foundation
- NIH HHS
- NCI NIH HHS
- Komen Foundation
- National Institutes of Health
Zitieren
APA
Krishnamoorthy, S., Morales, E., Ashmanskas, W. J., Werner, M. E., Matej, S., Raj, J., Vent, T. L., Choi, C. J., Maidment, A. D. A., Mankoff, D. A., Edmonds, C. E., Karp, J. S., & Surti, S. (2026). Design and imaging performance of BPET–DBT – a dedicated scanner for breast imaging. Medical Physics. https://doi.org/10.1002/mp.70651
Vancouver
Krishnamoorthy S, Morales E, Ashmanskas WJ, Werner ME, Matej S, Raj J, et al. Design and imaging performance of BPET–DBT – a dedicated scanner for breast imaging. Medical Physics. 2026. doi:10.1002/mp.70651
BibTeX
@article{Krishnamoorthy2026,
author = {Krishnamoorthy, Srilalan and Morales, Emmanuel and Ashmanskas, William J and Werner, Matthew E and Matej, Samuel and Raj, Juhi and Vent, Trevor L and Choi, Chloe J and Maidment, Andrew D A and Mankoff, David A and Edmonds, Christine E and Karp, Joel S and Surti, Suleman},
title = {Design and imaging performance of BPET–DBT – a dedicated scanner for breast imaging},
journal = {Medical Physics},
year = {2026},
doi = {10.1002/mp.70651}
}
EUVIMED