In‐line spectroscopy for iodine quantification in dynamic contrast‐enhanced dedicated breast CT
Medical Physics · 2025
DOI: 10.1002/mp.17497 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Liselot C Goris und 3 weiteren Autor:innen. Beteiligte Einrichtung: Technical Medicine Centre, University of Twente, Enschede, The Netherlands. Hauptthemen laut MEDLINE: Iodine, Contrast Media, Tomography, X-Ray Computed, Spectrum Analysis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program.
Abstract (Originalsprache)
Background
In 4D dynamic contrast‐enhanced dedicated breast computed tomography (4D DCE‐bCT), the functional properties of the breast will be characterized by monitoring the uptake and washout of iodine‐based contrast agents over time. This information could be valuable in breast cancer treatment. However, prior to clinical implementation, it is crucial to validate the quantitative estimates of iodine concentrations at each time point during acquisition.
Purpose
To develop an in‐line spectroscopy system capable of measuring iodine concentrations in a dynamic x‐ray breast phantom in real‐time.
Methods
Potassium iodide served as the contrast agent. The system was set‐up at both the entrance and exit of the phantom. It comprises a fiber‐coupled green LED and collimator, which together ensure that a parallel beam passes through the sample holder. Transmitted light is captured by a collimator on the opposite side and directed through a fiber optic cable to a photodetector for intensity measurement. The relationship between 13 iodine concentrations (0–6 mg I/mL) and light transmission was tested, and the system's repeatability and accuracy were determined.
Results
The system exhibited a strong correlation between iodine concentration and transmission values, achieving a root‐mean‐square error of 0.007. The repeated measurements had relative standard deviations of 0.04% and 0.1% for repeated water measurements at the phantom's entrance and exit, respectively. Furthermore, the accuracy measurements gave a mean error of fitting of 0.008 (± 0.07) mg I/mL.
Conclusion
The in‐line spectroscopy system can effectively monitor iodine concentrations in a dynamic breast phantom, providing a reliable method for quantitative validation of 4D DCE‐bCT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Technical Medicine Centre, University of Twente, Enschede, The Netherlands.
- Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
- Dutch Expert Centre for Screening (LRCB), Nijmegen, The Netherlands.
Förderung
- European Research Council (ERC) under the European Union's Horizon 2020 research and innovation program
Zitieren
APA
Goris, L. C., Mikerov, M., Pautasso, J. J., & Sechopoulos, I. (2025). In‐line spectroscopy for iodine quantification in dynamic contrast‐enhanced dedicated breast CT. Medical Physics. https://doi.org/10.1002/mp.17497
Vancouver
Goris LC, Mikerov M, Pautasso JJ, Sechopoulos I. In‐line spectroscopy for iodine quantification in dynamic contrast‐enhanced dedicated breast CT. Medical Physics. 2025. doi:10.1002/mp.17497
BibTeX
@article{Goris2025,
author = {Goris, Liselot C and Mikerov, Mikhail and Pautasso, Juan J and Sechopoulos, Ioannis},
title = {In‐line spectroscopy for iodine quantification in dynamic contrast‐enhanced dedicated breast CT},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.17497}
}
EUVIMED