Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices
Medical Physics · 2023
DOI: 10.1002/mp.16603 ↗At a glance
Journal article in Medical Physics, 2023. Written by Benson Yang and 2 co-authors. Institution: Sunnybrook Research Institute - Physical Sciences Platform, Toronto, ON, Canada. Major MEDLINE topics: Deep Brain Stimulation. 16 references.
Abstract (original language)
Background
Parallel radiofrequency transmission (pTx) remains a promising technology for addressing high‐field magnetic resonance imaging (MRI) challenges, particularly regarding the safety of patients with implanted deep brain stimulation (DBS) devices. Radiofrequency (RF) shim optimization methods utilizing pTx technology have shown the potential to minimize induced RF heating effects at the electrode tips of DBS devices at 3 T.
Purpose
Research pTx system implementations often involve the combination of custom and commercial hardware that are integrated onto an existing MRI system. As a result, system characterization is important to ensure implant‐friendly safe imaging conditions are satisfied for the operating range of the hardware.
Methods
Utilizing electromagnetic and thermal simulations, the impact of system uncertainty is studied for the proposed 4‐ and 8‐channel pTx system setup and its associated “safe mode” for DBS applications.
Results
Electromagnetic simulations indicated that instrumentation errors can affect the overall electric field strength experienced at the DBS lead tip, and a worst‐case system uncertainty analysis predicted temperature elevations of +1.5°C in the 4‐channel setup and +0.9°C in the 8‐channel setup.
Conclusions
In conclusion, system uncertainty can impact the precision of pTx RF inputs which in the worst‐case, may lead to an unsafe imaging scenario and the proposed 8‐channel setup may provide more robustness and thus, safer conditions for MRI of DBS patients.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Sunnybrook Research Institute - Physical Sciences Platform, Toronto, ON, Canada.
- Department of Electrical and Computer Engineering, McMaster University, Hamilton, ON, Canada.
- Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada.
Cite
APA
Yang, B., Chen, C. H., & Graham, S. J. (2023). Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices. Medical Physics. https://doi.org/10.1002/mp.16603
Vancouver
Yang B, Chen CH, Graham SJ. Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices. Medical Physics. 2023. doi:10.1002/mp.16603
BibTeX
@article{Yang2023,
author = {Yang, Benson and Chen, Chih-Hung and Graham, Simon J},
title = {Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices},
journal = {Medical Physics},
year = {2023},
doi = {10.1002/mp.16603}
}
EUVIMED