Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices

Benson Yang, Chih-Hung Chen, Simon J Graham

Medical Physics · 2023

DOI: 10.1002/mp.16603 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Benson Yang und 2 weiteren Autor:innen. Beteiligte Einrichtung: Sunnybrook Research Institute - Physical Sciences Platform, Toronto, ON, Canada. Hauptthemen laut MEDLINE: Deep Brain Stimulation. Die Arbeit nennt 16 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Parallel radiofrequency transmission (pTx) remains a promising technology for addressing high‐field magnetic resonance imaging (MRI) challenges, particularly regarding the safety of patients with implanted deep brain stimulation (DBS) devices. Radiofrequency (RF) shim optimization methods utilizing pTx technology have shown the potential to minimize induced RF heating effects at the electrode tips of DBS devices at 3 T.

Purpose

Research pTx system implementations often involve the combination of custom and commercial hardware that are integrated onto an existing MRI system. As a result, system characterization is important to ensure implant‐friendly safe imaging conditions are satisfied for the operating range of the hardware.

Methods

Utilizing electromagnetic and thermal simulations, the impact of system uncertainty is studied for the proposed 4‐ and 8‐channel pTx system setup and its associated “safe mode” for DBS applications.

Results

Electromagnetic simulations indicated that instrumentation errors can affect the overall electric field strength experienced at the DBS lead tip, and a worst‐case system uncertainty analysis predicted temperature elevations of +1.5°C in the 4‐channel setup and +0.9°C in the 8‐channel setup.

Conclusions

In conclusion, system uncertainty can impact the precision of pTx RF inputs which in the worst‐case, may lead to an unsafe imaging scenario and the proposed 8‐channel setup may provide more robustness and thus, safer conditions for MRI of DBS patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep Brain StimulationHumansUncertaintyMagnetic Resonance ImagingProstheses and ImplantsPhantoms, ImagingRadio Waves

Stichwörter der Autor:innen

MRI safetydeep brain stimulationinstrument uncertaintyparallel RF transmissionsystem error

Autor:innen und Institutionen

  • Sunnybrook Research Institute - Physical Sciences Platform, Toronto, ON, Canada.
  • Department of Electrical and Computer Engineering, McMaster University, Hamilton, ON, Canada.
  • Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada.

Zitieren

APA

Yang, B., Chen, C. H., & Graham, S. J. (2023). Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices. Medical Physics. https://doi.org/10.1002/mp.16603

Vancouver

Yang B, Chen CH, Graham SJ. Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices. Medical Physics. 2023. doi:10.1002/mp.16603

BibTeX
@article{Yang2023,
  author = {Yang, Benson and Chen, Chih-Hung and Graham, Simon J},
  title = {Technical note: System uncertainty on four‐ and eight‐channel parallel RF transmission for safe MRI of deep brain stimulation devices},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16603}
}
Kostenlos starten →