Journal article in Medical Physics, 2023. Written by Roman Vasyltsiv and 5 co-authors. Institution: Department of Radiology, Stony Brook University, Health Sciences Center L4-120, Stony Brook, New York, USA. Major MEDLINE topics: Brachytherapy. Cited 4 times so far (Crossref); 59 references.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

High dose rate (HDR) brachytherapy rapidly delivers dose to targets with steep dose gradients. This treatment method must adhere to prescribed treatment plans with high spatiotemporal accuracy and precision, as failure to do so may degrade clinical outcomes. One approach to achieving this goal is to develop imaging techniques to track HDR sources in vivo in reference to surrounding anatomy. This work investigates the feasibility of using an isocentric C‐arm x‐ray imager and tomosynthesis methods to track Ir‐192 HDR brachytherapy sources in vivo over time (4D).

Methods

A tomosynthesis imaging workflow was proposed and its achievable source detectability, localization accuracy, and spatiotemporal resolution were investigated in silico . An anthropomorphic female XCAT phantom was modified to include a vaginal cylinder applicator and Ir‐192 HDR source (0.5 × 0.5 × 5.0 mm 3 ), and the workflow was carried out using the MC‐GPU Monte Carlo image simulation platform. Source detectability was characterized using the reconstructed source signal‐difference‐to‐noise‐ratio (SDNR), localization accuracy by the absolute 3D error in its measured centroid location, and spatiotemporal resolution by the full‐width‐at‐half‐maximum (FWHM) of line profiles through the source in each spatial dimension considering a maximum C‐arm angular velocity of 30° per second. The dependence of these parameters on acquisition angular range ( θ tot = 0°–90°), number of views, angular increment between views (Δ θ = 0°–15°), and volumetric constraints imposed in reconstruction was evaluated. Organ voxel doses were tallied to derive the workflow's attributable effective dose.

Results

The HDR source was readily detected and its centroid was accurately localized with the proposed workflow and method (SDNR: 10–40, 3D error: 0–0.144 mm). Tradeoffs were demonstrated for various combinations of image acquisition parameters; namely, increasing the tomosynthesis acquisition angular range improved resolution in the depth‐encoded direction, for example from 2.5 mm to 1.2 mm between θ tot = 30 o and θ tot = 90 o , at the cost of increasing acquisition time from 1 to 3 s. The best‐performing acquisition parameters ( θ tot = 90 o , Δ θ = 1°) yielded no centroid localization error, and achieved submillimeter source resolution (0.57 × 1.21 × 5.04 mm 3 apparent source dimensions, FWHM). The total effective dose for the workflow was 263 µSv for its required pre‐treatment imaging component and 7.59 µSv per mid‐treatment acquisition thereafter, which is comparable to common diagnostic radiology exams.

Conclusions

A system and method for tracking HDR brachytherapy sources in vivo using C‐arm tomosynthesis was proposed and its performance investigated in silico . Tradeoffs in source conspicuity, localization accuracy, spatiotemporal resolution, and dose were determined. The results suggest this approach is feasible for localizing an Ir‐192 HDR source in vivo with submillimeter spatial resolution, 1–3 second temporal resolution and minimal additional dose burden.

Subjects

MeSH terms (MEDLINE)

BrachytherapyHumansFemaleRadiotherapy DosageX-RaysFeasibility StudiesPhantoms, ImagingMonte Carlo MethodIridium Radioisotopes

Author keywords

CBCTHDR brachytherapyIGRTMonte Carlosource trackingx-ray tomosynthesis

Substances

Iridium-192Iridium Radioisotopes

Authors and institutions

  • Department of Radiology, Stony Brook University, Health Sciences Center L4-120, Stony Brook, New York, USA.
  • Department of Radiation Oncology, Stony Brook University, Health Sciences Center L2, Stony Brook, New York, USA.

Cite

APA

Vasyltsiv, R., Qian, X., Xu, Z., Ryu, S., Zhao, W., & Howansky, A. (2023). Feasibility of 4D HDR brachytherapy source tracking using x‐ray tomosynthesis: Monte Carlo investigation. Medical Physics. https://doi.org/10.1002/mp.16579

Vancouver

Vasyltsiv R, Qian X, Xu Z, Ryu S, Zhao W, Howansky A. Feasibility of 4D HDR brachytherapy source tracking using x‐ray tomosynthesis: Monte Carlo investigation. Medical Physics. 2023. doi:10.1002/mp.16579

BibTeX
@article{Vasyltsiv2023,
  author = {Vasyltsiv, Roman and Qian, Xin and Xu, Zhigang and Ryu, Samuel and Zhao, Wei and Howansky, Adrian},
  title = {Feasibility of 4D HDR brachytherapy source tracking using x‐ray tomosynthesis: Monte Carlo investigation},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16579}
}
Start for free →