Development of a hydrated electron dosimeter for radiotherapy applications: A proof of concept

Julien Mégrourèche, Hamed Bekerat, Jingyi Bian, Alaina Bui, Jack Sankey, Lilian Childress, Shirin A Enger

Medical Physics · 2023

DOI: 10.1002/mp.16555 ↗

At a glance

Journal article in Medical Physics, 2023. Written by Julien Mégrourèche and 6 co-authors. Institution: Medical Physics Unit, Department of Oncology, McGill University, Montréal, Québec, Canada. Major MEDLINE topics: Electrons, Radiation Dosimeters. Cited 9 times so far (Crossref); 20 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: MEDTEQ+, Canada Research Chairs, Natural Sciences and Engineering Research Council of Canada.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Hydrated electrons, which are short‐lived products of radiolysis in water, increase the optical absorption of water, providing a pathway toward near‐tissue‐equivalent clinical radiation dosimeters. This has been demonstrated in high‐dose‐per‐pulse radiochemistry research, but, owing to the weak absorption signal, its application in existing low‐dose‐per‐pulse radiotherapy provided by clinical linear accelerators (linacs) has yet to be investigated.

Purpose

The aims of this study were to measure the optical absorption associated with hydrated electrons produced by clinical linacs and to assess the suitability of the technique for radiotherapy (⩽ 1 cGy per pulse) applications.

Methods

40 mW of 660‐nm laser light was sent five passes through deionized water contained in a 102 cm 3 glass‐walled cavity by using four broadband dielectric mirrors, two on each side of the cavity. The light was collected with a biased silicon photodetector. The water cavity was then irradiated by a Varian TrueBeam linac with both photon (10 MV FFF, 6 MV FFF, 6 MV) and electron beams (6 MeV) while monitoring the transmitted laser power for absorption transients. Radiochromic EBT3 film measurements were also performed for comparison.

Results

Examination of the absorbance profiles showed clear absorption changes in the water when radiation pulses were delivered. Both the amplitude and the decay time of the signal appeared consistent with the absorbed dose and the characteristics of the hydrated electrons. By using literature value for the hydrated electron radiation chemical yield (3.0±0.3), we inferred doses of 2.1±0.2 mGy (10 MV FFF), 1.3±0.1 mGy (6 MV FFF), 0.45±0.06 mGy (6 MV) for photons, and 0.47±0.05 mGy (6 MeV) for electrons, which differed from EBT3 film measurements by 0.6%, 0.8%, 10%, and 15.7%, respectively. The half‐life of the hydrated electrons in the solution was ∼ 24 s.

Conclusions

By measuring 660‐nm laser light transmitted through a cm‐scale, multi‐pass water cavity, we observed absorption transients consistent with hydrated electrons generated by clinical linac radiation. The agreement between our inferred dose and EBT3 film measurements suggests this proof‐of‐concept system represents a viable pathway toward tissue‐equivalent dosimeters for clinical radiotherapy applications.

Subjects

MeSH terms (MEDLINE)

ElectronsRadiation DosimetersPhotonsPhantoms, ImagingParticle AcceleratorsWaterRadiotherapy DosageRadiometry

Author keywords

hydrated electronradiation detectorradiolysis

Substances

Water

Authors and institutions

  • Medical Physics Unit, Department of Oncology, McGill University, Montréal, Québec, Canada.
  • Radiation Oncology Department, Jewish General Hospital, Montréal, Québec, Canada.
  • Department of Physics, McGill University, Montréal, Québec, Canada.
  • Research Institute of the McGill University Health Centre, Montréal, Québec, Canada.
  • Lady Davis Institute for Medical Research, Jewish General Hospital, Montréal, Québec, Canada.

Funding

  • MEDTEQ+
  • Canada Research Chairs
  • Natural Sciences and Engineering Research Council of Canada

Cite

APA

Mégrourèche, J., Bekerat, H., Bian, J., Bui, A., Sankey, J., Childress, L., & Enger, S. A. (2023). Development of a hydrated electron dosimeter for radiotherapy applications: A proof of concept. Medical Physics. https://doi.org/10.1002/mp.16555

Vancouver

Mégrourèche J, Bekerat H, Bian J, Bui A, Sankey J, Childress L, et al. Development of a hydrated electron dosimeter for radiotherapy applications: A proof of concept. Medical Physics. 2023. doi:10.1002/mp.16555

BibTeX
@article{Mgrourche2023,
  author = {Mégrourèche, Julien and Bekerat, Hamed and Bian, Jingyi and Bui, Alaina and Sankey, Jack and Childress, Lilian and Enger, Shirin A},
  title = {Development of a hydrated electron dosimeter for radiotherapy applications: A proof of concept},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16555}
}
Start for free →