Development and verification of an electron Monte Carlo engine for applications in intraoperative radiation therapy
Medical Physics · 2024
DOI: 10.1002/mp.17180 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Medical Physics, 2024. Verfasst von Luisa Rank und 6 weiteren Autor:innen. Beteiligte Einrichtung: Heidelberg Ion-Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany. Hauptthemen laut MEDLINE: Monte Carlo Method, Electrons. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: German Federal Ministry of Education and Research (BMBF).
Abstract (Originalsprache)
Background
In preparation of future clinical trials employing the Mobetron electron linear accelerator to deliver FLASH Intraoperative Radiation Therapy (IORT), the development of a Monte Carlo (MC)‐based framework for dose calculation was required.
Purpose
To extend and validate the in‐house developed fast MC dose engine MonteRay (MR) for future clinical applications in IORT.
Methods
MR is a CPU MC dose calculation engine written in C++ that is capable of simulating therapeutic proton, helium, and carbon ion beams. In this work, development steps are taken to include electrons and photons in MR are presented. To assess MRs accuracy, MR generated simulation results were compared against FLUKA predictions in water, in presence of heterogeneities as well as in an anthropomorphic phantom. Additionally, dosimetric data has been acquired to evaluate MRs accuracy in predicting dose‐distributions generated by the Mobetron accelerator. Runtimes of MR were evaluated against those of the general‐purpose MC code FLUKA on standard benchmark problems.
Results
MR generated dose distributions for electron beams incident on a water phantom match corresponding FLUKA calculated distributions within 2.3% with range values matching within 0.01 mm. In terms of dosimetric validation, differences between MR calculated and measured dose values were below 3% for almost all investigated positions within the water phantom. Gamma passing rate (1%/1 mm) for the scenarios with inhomogeneities and gamma passing rate (3%/2 mm) with the anthropomorphic phantom, were > 99.8% and 99.4%, respectively. The average dose differences between MR (FLUKA) and the measurements was 1.26% (1.09%). Deviations between MR and FLUKA were well within 1.5% for all investigated depths and 0.6% on average. In terms of runtime, MR achieved a speedup against reference FLUKA simulations of about 13 for 10 MeV electrons.
Conclusions
Validations against general purpose MC code FLUKA predictions and experimental dosimetric data have proven the validity of the physical models implemented in MR for IORT applications. Extending the work presented here, MR will be interfaced with external biophysical models to allow accurate FLASH biological dose predictions in IORT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Heidelberg Ion-Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
- Karlsruhe Institute of Technology (KIT), Faculty of Physics, Karlsruhe, Germany.
- Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
- Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg University Hospital (UKHD), Heidelberg Faculty of Medicine (MFHD) and German Cancer Research Center (DKFZ), He…
- Clinical Cooperation Unit Translational Radiation Oncology, German Cancer Consortium (DKTK) Core-Center Heidelberg, National Center for Tumor Diseases (NCT), Heidelberg University Hospital (UKHD) and German Cancer Resea…
- Division of Molecular and Translational Radiation Oncology, Heidelberg Faculty of Medicine (MFHD) and Department of Radiation Oncology, Heidelberg University Hospital (UKHD), Heidelberg, Germany.
Förderung
- German Federal Ministry of Education and Research (BMBF)
Zitieren
APA
Rank, L., Lysakovski, P., Major, G., Ferrari, A., Tessonnier, T., Debus, J., & Mairani, A. (2024). Development and verification of an electron Monte Carlo engine for applications in intraoperative radiation therapy. Medical Physics. https://doi.org/10.1002/mp.17180
Vancouver
Rank L, Lysakovski P, Major G, Ferrari A, Tessonnier T, Debus J, et al. Development and verification of an electron Monte Carlo engine for applications in intraoperative radiation therapy. Medical Physics. 2024. doi:10.1002/mp.17180