Characterization of a diode dosimeter for UHDR FLASH radiotherapy
Medical Physics · 2023-09 · 50 (9) · 5875-5883
DOI: 10.1002/mp.16474 ↗At a glance
Journal article in Medical Physics, published online on 2023-05-30. Written by Mahbubur Rahman and 10 co-authors. Institution: Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA. Major MEDLINE topics: Radiation Dosimeters, In Vivo Dosimetry. Cited 28 times so far (Crossref); 33 references. The full text is freely available in PubMed Central. Funded by: NCI NIH HHS, NIBIB NIH HHS.
Abstract (original language)
Background
Ultra‐high dose rate (UHDR) FLASH beams typically deliver dose at rates of >40 Gy/sec. Characterization of these beams with respect to dose, mean dose rate, and dose per pulse requires dosimeters which exhibit high temporal resolution and fast readout capabilities.
Purpose
A diode EDGE Detector with a newly designed electrometer has been characterized for use in an UHDR electron beam and demonstrated appropriateness for UHDR FLASH radiotherapy dosimetry.
Methods
Dose linearity, mean dose rate, and dose per pulse dependencies of the EDGE Detector were quantified and compared with dosimeters including a W1 scintillator detector, radiochromic film, and ionization chamber that were irradiated with a 10 MeV UHDR beam. The dose, dose rate, and dose per pulse were controlled via an in‐house developed scintillation‐based feedback mechanism, repetition rate of the linear accelerator, and source‐to‐surface distance, respectively. Depth‐dose profiles and temporal profiles at individual pulse resolution were compared to the film and scintillation measurements, respectively. The radiation‐induced change in response sensitivity was quantified via irradiation of ∼5kGy.
Results
The EDGE Detector agreed with film measurements in the measured range with varying dose (up to 70 Gy), dose rate (nearly 200 Gy/s), and dose per pulse (up to 0.63 Gy/pulse) on average to within 2%, 5%, and 1%, respectively. The detector also agreed with W1 scintillation detector on average to within 2% for dose per pulse (up to 0.78 Gy/pulse). The EDGE Detector signal was proportional to ion chamber (IC) measured dose, and mean dose rate in the bremsstrahlung tail to within 0.4% and 0.2% respectively. The EDGE Detector measured percent depth dose (PDD) agreed with film to within 3% and per pulse output agreed with W1 scintillator to within −6% to +5%. The radiation‐induced response decrease was 0.4% per kGy.
Conclusions
The EDGE Detector demonstrated dose linearity, mean dose rate independence, and dose per pulse independence for UHDR electron beams. It can quantify the beam spatially, and temporally at sub millisecond resolution. It's robustness and individual pulse detectability of treatment deliveries can potentially lead to its implementation for in vivo FLASH dosimetry, and dose monitoring.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
- UT Southwestern Medical Center, Dallas, Texas, USA.
- Sun Nuclear Inc., Melbourne, Florida, USA.
- Stanford University, Stanford, California, USA.
- Department of Medicine, Radiation Oncology, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.
- Dartmouth Cancer Center, Dartmouth-Hitchcock Medical Center, Lebanon, New Hampshire, USA.
Funding
- NCI NIH HHS
- NIBIB NIH HHS
Cite
APA
Rahman, M., Kozelka, J., Hildreth, J., Schönfeld, A., Sloop, A. M., Ashraf, M. R., Bruza, P., Gladstone, D. J., Pogue, B. W., Simon, W. E., & Zhang, R. (2023). Characterization of a diode dosimeter for UHDR FLASH radiotherapy. Medical Physics, 50(9), 5875-5883. https://doi.org/10.1002/mp.16474
Vancouver
Rahman M, Kozelka J, Hildreth J, Schönfeld A, Sloop AM, Ashraf MR, et al. Characterization of a diode dosimeter for UHDR FLASH radiotherapy. Medical Physics. 2023;50(9):5875-5883. doi:10.1002/mp.16474
BibTeX
@article{Rahman2023,
author = {Rahman, Mahbubur and Kozelka, Jakub and Hildreth, Jeff and Schönfeld, Andreas and Sloop, Austin M and Ashraf, M Ramish and Bruza, Petr and Gladstone, David J and Pogue, Brian W and Simon, William E and Zhang, Rongxiao},
title = {Characterization of a diode dosimeter for UHDR FLASH radiotherapy},
journal = {Medical Physics},
year = {2023},
volume = {50},
number = {9},
pages = {5875--5883},
doi = {10.1002/mp.16474}
}
EUVIMED