EUVIMEDEUVIMED entdecken
Crossref

Technical note: Comprehensive evaluations of gantry and couch rotation isocentricities for implementing proton stereotactic radiosurgery

Jiajian Shen, Daniel G Robertson, Martin Bues, Konstantin Shipulin, Wei Liu, Joshua Stoker, Jonathan B Ashman, Pedro Lara, Sameer R Keole, William Wong, Sujay A Vora

Medical Physics · 2023

DOI: 10.1002/mp.16382 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Jiajian Shen und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA. Hauptthemen laut MEDLINE: Radiosurgery. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 15 Literaturangaben. Gefördert von: The Kemper Marley Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Mechanical accuracy should be verified before implementing a proton stereotactic radiosurgery (SRS) program. Linear accelerator (Linac)‐based SRS systems often use electronic portal imaging devices (EPIDs) to verify beam isocentricity. Because proton therapy systems do not have EPID, beam isocentricity tests of proton SRS may still rely on films, which are not efficient.

Purpose

To validate that our proton SRS system meets mechanical precision requirements and to present an efficient method to evaluate the couch and gantry's rotational isocentricity for our proton SRS system.

Methods

A dedicated applicator to hold brass aperture for proton SRS system was designed. The mechanical precision of the system was tested using a metal ball and film for 11 combinations of gantry and couch angles. A more efficient quality assurance (QA) procedure was developed, which used a scintillator device to replace the film. The couch rotational isocentricity tests were performed using orthogonal kV x‐rays with the couch rotated isocentrically to five positions (0°, 315°, 270°, 225°, and 180°). At each couch position, the distance between the metal ball in kV images and the imaging isocenter was measured. The gantry isocentricity tests were performed using a cone‐shaped scintillator and proton beams at five gantry angles (0°, 45°, 90°, 135°, and 180°), and the isocenter position and the distance of each beam path to the isocenter were obtained. Daily QA procedure was performed for 1 month to test the robustness and reproducibility of the procedure.

Results

The gantry and couch rotational isocentricity exhibited sub‐mm precision, with most measurements within ±0.5 mm. The 1‐month QA results showed that the procedure was robust and highly reproducible to within ±0.2 mm. The gantry isocentricity test using the cone‐shaped scintillator was accurate and sensitive to variations of ±0.2 mm. The QA procedure was efficient enough to be completed within 30 min. The 1‐month isocentricity position variations were within 0.5 mm, which demonstrating that the overall proton SRS system was stable and precise.

Conclusion

The proton SRS Winston–Lutz QA procedure using a cone‐shaped scintillator was efficient and robust. We were able to verify radiation delivery could be performed with sub‐mm mechanical precision.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryProtonsRotationReproducibility of ResultsDiagnostic ImagingParticle AcceleratorsPhantoms, Imaging

Stichwörter der Autor:innen

Winston-Lutz testisocentricity coincidenceproton SRSquality assurancerobotic couch

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.

Förderung

  • The Kemper Marley Foundation

Zitieren

APA

Shen, J., Robertson, D. G., Bues, M., Shipulin, K., Liu, W., Stoker, J., Ashman, J. B., Lara, P., Keole, S. R., Wong, W., & Vora, S. A. (2023). Technical note: Comprehensive evaluations of gantry and couch rotation isocentricities for implementing proton stereotactic radiosurgery. Medical Physics. https://doi.org/10.1002/mp.16382

Vancouver

Shen J, Robertson DG, Bues M, Shipulin K, Liu W, Stoker J, et al. Technical note: Comprehensive evaluations of gantry and couch rotation isocentricities for implementing proton stereotactic radiosurgery. Medical Physics. 2023. doi:10.1002/mp.16382

BibTeX
@article{Shen2023,
  author = {Shen, Jiajian and Robertson, Daniel G and Bues, Martin and Shipulin, Konstantin and Liu, Wei and Stoker, Joshua and Ashman, Jonathan B and Lara, Pedro and Keole, Sameer R and Wong, William and Vora, Sujay A},
  title = {Technical note: Comprehensive evaluations of gantry and couch rotation isocentricities for implementing proton stereotactic radiosurgery},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16382}
}
Kostenlos starten →