EUVIMEDOpdag EUVIMED
Crossref

Image‐based scatter correction for cone‐beam CT using flip swin transformer U‐shape network

Xueren Zhang, Yangkang Jiang, Chen Luo, Dengwang Li, Tianye Niu, Gang Yu

Medical Physics · 2023-08 · 50 (8) · 5002-5019

DOI: 10.1002/mp.16277 ↗

At a glance

Journal article in Medical Physics, published online on 2023-02-14. Written by Xueren Zhang and 5 co-authors. Institution: Shandong Key Laboratory of Medical Physics and Image Processing, Shandong Institute of Industrial Technology for Health Sciences and Precision Medici…. Major MEDLINE topics: Algorithms, Neural Networks, Computer. Cited 14 times so far (Crossref); 51 references. Funded by: National Nature Science Foundation of China, Natural Science Foundation of Shandong Province, National Key R&D Program of China.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Cone beam computed tomography (CBCT) plays an increasingly important role in image‐guided radiation therapy. However, the image quality of CBCT is severely degraded by excessive scatter contamination, especially in the abdominal region, hindering its further applications in radiation therapy.

Purpose

To restore low‐quality CBCT images contaminated by scatter signals, a scatter correction algorithm combining the advantages of convolutional neural networks (CNN) and Swin Transformer is proposed.

Methods

In this paper a scatter correction model for CBCT image, the Flip Swin Transformer U‐shape network (FSTUNet) model, is proposed. In this model, the advantages of CNN in texture detail and Swin Transformer in global correlation are used to accurately extract shallow and deep features, respectively. Instead of using the original Swin Transformer tandem structure, we build the Flip Swin Transformer Block to achieve a more powerful inter‐window association extraction. The validity and clinical relevance of the method is demonstrated through extensive experiments on a Monte Carlo (MC) simulation dataset and frequency split dataset generated by a validated method, respectively.

Result

Experimental results on the MC simulated dataset show that the root mean square error of images corrected by the method is reduced from over 100 HU to about 7 HU. Both the structural similarity index measure (SSIM) and the universal quality index (UQI) are close to 1. Experimental results on the frequency split dataset demonstrate that the method not only corrects shading artifacts but also exhibits a high degree of structural consistency. In addition, comparison experiments show that FSTUNet outperforms UNet, Deep Residual Convolutional Neural Network (DRCNN), DSENet, Pix2pixGAN, and 3DUnet methods in both qualitative and quantitative metrics.

Conclusions

Accurately capturing the features at different levels is greatly beneficial for reconstructing high‐quality scatter‐free images. The proposed FSTUNet method is an effective solution to CBCT scatter correction and has the potential to improve the accuracy of CBCT image‐guided radiation therapy.

Subjects

MeSH terms (MEDLINE)

AlgorithmsNeural Networks, ComputerScattering, RadiationPhantoms, ImagingCone-Beam Computed Tomography

Author keywords

CBCTCNNscatter correctionswin transformer

Authors and institutions

  • Shandong Key Laboratory of Medical Physics and Image Processing, Shandong Institute of Industrial Technology for Health Sciences and Precision Medicine, School of Physics and Electronics, Shandong Normal University, Jin…
  • Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, Institute of Translational Medicine, Zhejiang University, Hangzhou, Zhejiang, China.
  • Shenzhen Bay Laboratory, Shenzhen, China.
  • School of Automation, Zhejiang Institute of Mechanical & Electrical Engineering, Hangzhou, China.

Funding

  • National Nature Science Foundation of China
  • Natural Science Foundation of Shandong Province
  • National Key R&D Program of China

Cite

APA

Zhang, X., Jiang, Y., Luo, C., Li, D., Niu, T., & Yu, G. (2023). Image‐based scatter correction for cone‐beam CT using flip swin transformer U‐shape network. Medical Physics, 50(8), 5002-5019. https://doi.org/10.1002/mp.16277

Vancouver

Zhang X, Jiang Y, Luo C, Li D, Niu T, Yu G. Image‐based scatter correction for cone‐beam CT using flip swin transformer U‐shape network. Medical Physics. 2023;50(8):5002-5019. doi:10.1002/mp.16277

BibTeX
@article{Zhang2023,
  author = {Zhang, Xueren and Jiang, Yangkang and Luo, Chen and Li, Dengwang and Niu, Tianye and Yu, Gang},
  title = {Image‐based scatter correction for cone‐beam CT using flip swin transformer U‐shape network},
  journal = {Medical Physics},
  year = {2023},
  volume = {50},
  number = {8},
  pages = {5002--5019},
  doi = {10.1002/mp.16277}
}
Start for free →