Effect of Phase Encoding Direction on Image Quality in Single‐Shot EPI Diffusion‐Weighted Imaging of the Breast
Journal of Magnetic Resonance Imaging · 2024
DOI: 10.1002/jmri.29304 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2024. Verfasst von Ana E Rodríguez-Soto und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California San Diego, La Jolla, California, USA. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Artifacts, Phantoms, Imaging, Breast. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 37 Literaturangaben. Gefördert von: GE Healthcare, Krueger v. Wyeth Cy Pres Settlement Funds.
Abstract (Originalsprache)
Background
In breast diffusion‐weighted imaging (DWI), distortion and physiologic artifacts affect clinical interpretation. Image quality can be optimized by addressing the effect of phase encoding (PE) direction on these artifacts.
Purpose
To compare distortion artifacts in breast DWI acquired with different PE directions and polarities, and to discuss their clinical implications. Study Type Prospective. Population Eleven healthy volunteers (median age: 47 years old; range: 22–74 years old) and a breast phantom. Field Strength/Sequence Single‐shot echo planar DWI and three‐dimensional fast gradient echo sequences at 3 T. Assessment All DWI data were acquired with left–right, right–left, posterior–anterior, and anterior–posterior PE directions. In phantom data, displacement magnitude was evaluated by comparing the location of landmarks in anatomical and DWI images. Three breast radiologists (5, 17, and 23 years of experience) assessed the presence or absence of physiologic artifacts in volunteers' DWI datasets and indicated their PE‐direction preference. Statistical Tests Analysis of variance with post‐hoc tests were used to assess differences in displacement magnitude across DWI datasets and observers. A binomial test and a chi‐squared test were used to evaluate if each in vivo DWI dataset had an equal probability (25%) of being preferred by radiologists. Inter‐reader agreement was evaluated using Gwet's AC1 agreement coefficient. A P ‐value
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University of California San Diego, La Jolla, California, USA.
- Herbert Wertheim School of Public Health and Human Longevity Science, University of California San Diego, La Jolla, California, USA.
ORCID-Profile: Ana E Rodríguez-Soto, Stephane Loubrie, Alexandra Schlein, Haydee Ojeda-Fournier, Rebecca Rakow-Penner
Förderung
- GE Healthcare
- Krueger v. Wyeth Cy Pres Settlement Funds
Zitieren
APA
Rodríguez-Soto, A. E., Zou, J., Loubrie, S., Ebrahimi, S., Jordan, S., Schlein, A., Lim, V., Ojeda-Fournier, H., & Rakow-Penner, R. (2024). Effect of Phase Encoding Direction on Image Quality in Single‐Shot EPI Diffusion‐Weighted Imaging of the Breast. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29304
Vancouver
Rodríguez-Soto AE, Zou J, Loubrie S, Ebrahimi S, Jordan S, Schlein A, et al. Effect of Phase Encoding Direction on Image Quality in Single‐Shot EPI Diffusion‐Weighted Imaging of the Breast. Journal of Magnetic Resonance Imaging. 2024. doi:10.1002/jmri.29304
BibTeX
@article{RodrguezSoto2024,
author = {Rodríguez-Soto, Ana E and Zou, Jingjing and Loubrie, Stephane and Ebrahimi, Sheida and Jordan, Stephan and Schlein, Alexandra and Lim, Vivian and Ojeda-Fournier, Haydee and Rakow-Penner, Rebecca},
title = {Effect of Phase Encoding Direction on Image Quality in Single‐Shot EPI Diffusion‐Weighted Imaging of the Breast},
journal = {Journal of Magnetic Resonance Imaging},
year = {2024},
doi = {10.1002/jmri.29304}
}
EUVIMED