Analysis of Flexural Strength and Brittleness of a Polyjet 3D‐Printed Denture Base Polymer

Gregory W Bennett iD, Alex Kohler

Clinical and Experimental Dental Research · 2025

DOI: 10.1002/cre2.70228 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical and Experimental Dental Research, 2025. Verfasst von Gregory W Bennett und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Adult Restorative Dentistry, University of Nebraska Medical Center College of Dentistry, Lincoln, Nebraska, USA. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Flexural Strength, Denture Bases, Polymers. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study sought to understand the effects of print orientation on a novel 3D‐printing technology recently made available for the fabrication of dentures. Material and Methods A total of 90 experimental samples (30 each at 0, 45, and 90 degrees) were printed using denture resin on a polyjet printer. The milled control samples ( n = 10) were milled on a 5‐axis dental laboratory mill from a denture base disc. 96 samples were tested for flexural strength and extension at break using a 3‐point bend test on a universal testing machine. The data were analyzed using statistical software and evaluated for statistical significance using general linear models with a Dunnett test with α = 0.05.

Results

The results showed that the flexural strength and extension at break were both affected by print orientation with the 90‐degree orientation being the lowest performing group for both properties. The 0‐degree and 45‐degree orientations both exceeded the ISO minimum flexural strength of 65 MPa, with mean flexural strengths of 88.18 and 73.53 MPa. The 90‐degree group mean was well below the standard at 28.12 MPa. The milled sample group mean was 65.18 MPa. Extension at break showed similar results with less variation in the printed groups. Mean extension at break of the milled samples group was 15.05 mm, the 0‐degree was 5.99 mm, 45‐degree was 4.72 mm, and 90‐degree was 1.92 mm.

Conclusion

Flexural strength and extension at break were influenced by print orientation. A 0‐degree, or horizontal, print orientation yielded the highest values for flexural strength. The milled samples had significantly higher values for extension at break, but similar values for flexural strength.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Printing, Three-DimensionalFlexural StrengthDenture BasesPolymersMaterials TestingDental Stress AnalysisHumans

Stichwörter der Autor:innen

denture baseflexural strengthmaterial jettingprinting

Substanzen

Polymers

Autor:innen und Institutionen

  • Department of Adult Restorative Dentistry, University of Nebraska Medical Center College of Dentistry, Lincoln, Nebraska, USA.
  • Student Nebraska Wesleyan University, Lincoln, Nebraska, USA.

ORCID-Profile: Gregory W Bennett

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Bennett, G. W., & Kohler, A. (2025). Analysis of Flexural Strength and Brittleness of a Polyjet 3D‐Printed Denture Base Polymer. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70228

Vancouver

Bennett GW, Kohler A. Analysis of Flexural Strength and Brittleness of a Polyjet 3D‐Printed Denture Base Polymer. Clinical and Experimental Dental Research. 2025. doi:10.1002/cre2.70228

BibTeX
@article{Bennett2025,
  author = {Bennett, Gregory W and Kohler, Alex},
  title = {Analysis of Flexural Strength and Brittleness of a Polyjet 3D‐Printed Denture Base Polymer},
  journal = {Clinical and Experimental Dental Research},
  year = {2025},
  doi = {10.1002/cre2.70228}
}
Kostenlos starten →