The Sequence of Post‐Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound
Catheterization and Cardiovascular Interventions · 2026-07-23
DOI: 10.1002/ccd.70774 ↗At a glance
Journal article in Catheterization and Cardiovascular Interventions, published online on 2026-07-23. Written by Dou Bo-Xu and 2 co-authors. Institution: Department of Cardiovascular Medicine, First Affiliated Hospital of Harbin Medical University, Harbin, China. Major MEDLINE topics: Ultrasonography, Interventional, Stents, Coronary Artery Disease, Angioplasty, Balloon, Coronary. 23 references.
Abstract (original language)
Background
Longitudinal stent deformation (LSD) is a potentially serious complication of percutaneous coronary intervention (PCI). Although post‐dilation is a known risk factor for LSD, the influence of procedural factors such as dilation sequence remains poorly understood.
Aims
This study aimed to identify procedural factors associated with LSD, with a focus on post‐dilation technique.
Methods
In 529 patients (533 vessels) undergoing intravascular ultrasound (IVUS)‐guided PCI, stent length was measured after post‐dilation. Patients were categorized based on length change: elongation (> 2 mm or > 10% of nominal stent length; n = 71), unchanged (±2 mm or < 10% of nominal stent length; n = 457), and shortening (> 2 mm or > 10% of nominal stent length; n = 1, likely attributable to measurement error and not statistically significant). Procedural and lesion characteristics were analyzed.
Results
In univariate analysis, post‐dilation sequence ( p < 0.01) and balloon length ( p < 0.05) differed between groups. In multivariate logistic regression, a distal‐to‐proximal dilation sequence (OR 3.13, 95% CI 1.2–10.81) and higher LVEF (OR 1.07, 95% CI 1.02–1.12) were independently associated with stent elongation. Stent placement at ostium was protective (OR 0.45, 95% CI 0.16–0.96). In the elongation group, positive correlations were observed between elongation magnitude and lesion length, stent length, and dilation sequence (all p < 0.05).
Conclusion
The post‐dilation sequence significantly influences LSD. Initiating dilation from the proximal stent segment—known as the proximal optimization technique (POT)—may reduce the risk of stent deformation.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiovascular Medicine, First Affiliated Hospital of Harbin Medical University, Harbin, China.
ORCID profiles: Dou Bo-Xu
Cite
APA
Bo-Xu, D., Zi-Guang, S., & Ji-Yi, Z. (2026). The Sequence of Post‐Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70774
Vancouver
Bo-Xu D, Zi-Guang S, Ji-Yi Z. The Sequence of Post‐Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70774
BibTeX
@article{BoXu2026,
author = {Bo-Xu, Dou and Zi-Guang, Song and Ji-Yi, Zhao},
title = {The Sequence of Post‐Dilation Matters: Proximal Optimization Technique (POT) Reduces Longitudinal Stent Deformation Assessed by Intravascular Ultrasound},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70774}
}
EUVIMED