Index of Device Efficiency: IVUS‐Based Metric to Quantify Cutting Balloon Performance in Calcified Coronary Lesions
Catheterization and Cardiovascular Interventions · 2026-06 · 107 (7) · 2335-2344
DOI: 10.1002/ccd.70597 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 07.04.2026. Verfasst von Claudiu Ungureanu und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Hôpital de Jolimont, La Louvrière, Belgium. Hauptthemen laut MEDLINE: Angioplasty, Balloon, Coronary, Cardiac Catheters, Coronary Artery Disease, Coronary Vessels. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
Calcified coronary artery disease complicates percutaneous coronary intervention (PCI), yet no standardized parameter quantifies the efficiency of plaque‐modifying devices.
Aims
To introduce and evaluate the Index of Device Efficiency (IDE), a novel intravascular ultrasound (IVUS)‐derived metric quantifying the mechanical efficiency of cutting balloon (CB) angioplasty during the RODIN‐CUT technique.
Methods
In this multicenter retrospective study, 50 heavily calcified lesions (47 patients) underwent 270 IVUS pullbacks. The IDE was defined as the ratio of maximal luminal diameter achieved to the nominal CB diameter ×100. Stepwise IDE dynamics were assessed after sequential high‐pressure CB inflations. A plateau phase was defined as ≤ 0.3 mm further luminal diameter gain or ≤ 0.5 mm² area gain between inflations. The relationship between IDE ≥ 100% and post‐stent expansion was evaluated. Safety outcomes were systematically recorded.
Results
Mean age was 75 ± 8 years; 46% of lesions were in the left anterior descending artery and 30% in the left main bifurcation. Minimal lumen area increased from 4.0 ± 1.5 mm² at baseline to 5.8 ± 2.1 mm² after one inflation and to 8.8 ± 2.6 mm² after four inflations ( p < 0.001). IDE ≥ 100% was reached in 40% of lesions after one inflation versus 92% after four, with no further gain thereafter (plateau). Lesions with IDE ≥ 100% more frequently achieved optimal stent expansion (82% vs. 10%, p = 0.007). No major procedural complications occurred.
Conclusions
IDE is a simple, reproducible IVUS‐derived metric that quantifies CB efficiency, identifies a reproducible plateau after four inflations, and predicts optimal stent expansion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Hôpital de Jolimont, La Louvrière, Belgium.
- University Hospital Basel, Basel, Switzerland.
- Department of Cardiology, Clinique St. Joseph-Groupe Vivalia, Arlon, Belgium.
- Cardiology Department, Ares Monza Hospital, Bucharest, Romania.
- Department of Cardiovascular Sciences, Campus Bio-medico University Hospital, Rome, Italy.
- Cardiology Department, "Niculae Stancioiu" Heart Institute, Cluj-Napoca, Romania.
ORCID-Profile: Claudiu Ungureanu, Gregor Leibundgut, Giuseppe Colletti, Agostino Spano, Gabriele Gasparini, Pierfrancesco Agostoni, Alexandru Achim, Adrien Jossart, Enrico Poletti, Antonio Mangieri, Emmanouil S Brilakis
Zitieren
APA
Ungureanu, C., Leibundgut, G., Colletti, G., Dumitrascu, S., Spano, A., Cocoi, M., Avran, A., Gasparini, G., Agostoni, P., Spratt, J., Achim, A., Arain, S., Jossart, A., Boutaleb, A. M., Poletti, E., Mangieri, A., Brilakis, E. S., & Boukhris, M. (2026). Index of Device Efficiency: IVUS‐Based Metric to Quantify Cutting Balloon Performance in Calcified Coronary Lesions. Catheterization and Cardiovascular Interventions, 107(7), 2335-2344. https://doi.org/10.1002/ccd.70597
Vancouver
Ungureanu C, Leibundgut G, Colletti G, Dumitrascu S, Spano A, Cocoi M, et al. Index of Device Efficiency: IVUS‐Based Metric to Quantify Cutting Balloon Performance in Calcified Coronary Lesions. Catheterization and Cardiovascular Interventions. 2026;107(7):2335-2344. doi:10.1002/ccd.70597
BibTeX
@article{Ungureanu2026,
author = {Ungureanu, Claudiu and Leibundgut, Gregor and Colletti, Giuseppe and Dumitrascu, Silviu and Spano, Agostino and Cocoi, Mihai and Avran, Alexandre and Gasparini, Gabriele and Agostoni, Pierfrancesco and Spratt, James and Achim, Alexandru and Arain, Salman and Jossart, Adrien and Boutaleb, Amine Mamoun and Poletti, Enrico and Mangieri, Antonio and Brilakis, Emmanouil S and Boukhris, Marouane},
title = {Index of Device Efficiency: IVUS‐Based Metric to Quantify Cutting Balloon Performance in Calcified Coronary Lesions},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
volume = {107},
number = {7},
pages = {2335--2344},
doi = {10.1002/ccd.70597}
}
EUVIMED