Journal article in Catheterization and Cardiovascular Interventions, 2026. Written by Jing Yang and 8 co-authors. Institution: Department of Cardiology, Beijing Luhe Hospital, Affiliated to Capital Medical University, Beijing, China. Major MEDLINE topics: Coronary Artery Disease, Robotics, Percutaneous Coronary Intervention, Coronary Stenosis. Cited 2 times so far (Crossref); 21 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: National Natural Science Foundation of China.
Abstract (original language)
Background
International robotic‐assisted PCI platforms such as CorPath 200 and CorPath GRX have demonstrated procedural safety and efficacy but are limited by the absence of real‐time tactile feedback and cumbersome device exchange. The ETcath200, the first domestically developed robotic PCI system in China, introduces guidewire force‐sensing to provide real‐time visual resistance feedback that partially compensates for the lack of direct tactile sensation, along with streamlined device exchange and an intuitive joystick–roller interface.
Aims
This study aimed to evaluate the feasibility, safety, and procedural performance of the ETcath200 system in patients with complex coronary lesions.
Methods
In this prospective, single‐center, single‐arm trial, 20 patients with AHA/ACC type B2 or C complex coronary lesions underwent PCI using the ETcath200 system. Technical success was defined as full robotic completion without any unplanned manual input (including partial manual assistance or manual conversion); clinical success as residual stenosis < 30% with TIMI 3 flow and no in‐hospital major adverse cardiac events (MACE). Patients were followed for 30 days.
Results
All 20 patients (mean age 63.2 ± 12.0 years; 80% male) had unstable angina and complex coronary anatomy (79.2% type C lesions; 50% bifurcation, 65% diffuse). Most procedures were performed via radial access (80%), and the majority utilized drug‐eluting stents (65%). Technical success was achieved in 16 cases (80%), while four cases (20%) required manual conversion due to severe tortuosity or calcification ( n = 3) or guidewire‐induced dissection ( n = 1; managed successfully). All patients achieved clinical success; no device failures, MACE, or serious adverse events occurred at 30 days. Mean procedure duration was 61.0 ± 23.3 min; radiation dose was 0.07 ± 0.04 mSv.
Conclusions
The ETcath200 robotic PCI system demonstrated feasibility, safety, and reliable performance in complex coronary lesions, achieving high clinical success and excellent short‐term outcomes. Manual conversion was required only in anatomically challenging cases. These data support further multicenter evaluation of ETcath200 in complex coronary interventions.
Subjects
MeSH terms (MEDLINE)
Authors and institutions
- Department of Cardiology, Beijing Luhe Hospital, Affiliated to Capital Medical University, Beijing, China.
- Beijing WeMed Medical Technology Co. Ltd., Beijing, China.
- Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China.
ORCID profiles: Guangyao Zhai
Funding
- National Natural Science Foundation of China
Cite
APA
Yang, J., Fang, Y., Yan, Y., Liu, Z., Xie, J., Zhang, G., Zhang, Z., Sun, T., & Zhai, G. (2026). Feasibility and Safety of the ETcath200 Robotic PCI System for Complex Coronary Lesions: A Prospective Single‐Center Study. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.70488
Vancouver
Yang J, Fang Y, Yan Y, Liu Z, Xie J, Zhang G, et al. Feasibility and Safety of the ETcath200 Robotic PCI System for Complex Coronary Lesions: A Prospective Single‐Center Study. Catheterization and Cardiovascular Interventions. 2026. doi:10.1002/ccd.70488
BibTeX
@article{Yang2026,
author = {Yang, Jing and Fang, Yuqi and Yan, Yizhu and Liu, Zijing and Xie, Jing and Zhang, Ge and Zhang, Zhicheng and Sun, Tienan and Zhai, Guangyao},
title = {Feasibility and Safety of the ETcath200 Robotic PCI System for Complex Coronary Lesions: A Prospective Single‐Center Study},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
doi = {10.1002/ccd.70488}
}
EUVIMED