Commissural Misalignment Following Valve‐in‐Valve Transcatheter Aortic Valve Implantation
Catheterization and Cardiovascular Interventions · 2026-03 · 107 (4) · 1215-1224
DOI: 10.1002/ccd.70460 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 07.01.2026. Verfasst von Matthias Raschpichler und 13 weiteren Autor:innen. Beteiligte Einrichtung: University Hospital of Cardiac Surgery, Heart Center Leipzig at Leipzig University, Leipzig, Germany. Hauptthemen laut MEDLINE: Aortic Valve Stenosis, Aortic Valve, Transcatheter Aortic Valve Replacement, Prosthesis Failure. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Data on commissural misalignment (CMA) during valve‐in‐valve transcatheter aortic valve implantation (ViV‐TAVI) for valve failure after surgical aortic valve replacement (SAVR) is scarce.
Aims
To study the impact of CMA on clinical and hemodynamic outcomes following ViV‐TAVI for failed SAVR.
Methods
Data of patients who underwent ViV‐TAVI for failed SAVR valves at two institutions (Heart Center Leipzig at Leipzig University, Leipzig, Germany; Cedars‐Sinai Smidt Heart Institute, Los Angeles, USA) were retrospectively collected and compared regarding the existence of CMA. Outcomes of interest included Valve Academic Research Consortium (VARC)‐3‐based clinical endpoints, Computed Tomography‐based hypoattenuated leaflet thickening (HALT), hemodynamic outcome, and mid‐term all‐cause mortality.
Results
Of the 687 patients who underwent ViV‐TAVI, post‐procedural CT of sufficient quality to measure commissural alignment was available in 180 patients (47.8% females, mean age 78.0 years). Self‐expanding TAVI valves were used in 49.2% of the cases. CMA was found in 35 individuals (19.4%) and was associated with reduced VARC‐3‐based technical success (80% vs. 93%, p = 0.03), driven by increased rates of coronary obstruction (17.1% vs. 5.6%, p = 0.03). The incidence of HALT was 31.7%, without differences in patients with and without CMA. VARC‐3‐based device success was 62.9%, driven by an incidence of residual mean valve gradient (≥ 20 mmHg) of 35.4%, without differences between groups. Midterm survival after a mean of 747 days was 84.6% and comparable between groups.
Conclusions
For patients undergoing ViV‐TAVI for failed SAVR, technical success is reduced for cases of CMA due to higher odds of coronary obstruction, without differences in mid‐term survival.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University Hospital of Cardiac Surgery, Heart Center Leipzig at Leipzig University, Leipzig, Germany.
- Department of Internal Medicine/Cardiology, Heart Center Leipzig at Leipzig University, Leipzig, Germany.
- Cedars-Sinai Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.
ORCID-Profile: Matthias Raschpichler, Johannes Rotta Detto Loria, Holger Thiele, Michael A Borger, Mohamed Abdel-Wahab
Zitieren
APA
Raschpichler, M., Loria, J. R. D., Patel, V., Noack, T., Nüsser, K., Dumpies, O., Patel, D., Nagasaka, T., Madaan, P., Gupta, A., Thiele, H., Borger, M. A., Makkar, R., & Abdel-Wahab, M. (2026). Commissural Misalignment Following Valve‐in‐Valve Transcatheter Aortic Valve Implantation. Catheterization and Cardiovascular Interventions, 107(4), 1215-1224. https://doi.org/10.1002/ccd.70460
Vancouver
Raschpichler M, Loria JRD, Patel V, Noack T, Nüsser K, Dumpies O, et al. Commissural Misalignment Following Valve‐in‐Valve Transcatheter Aortic Valve Implantation. Catheterization and Cardiovascular Interventions. 2026;107(4):1215-1224. doi:10.1002/ccd.70460
BibTeX
@article{Raschpichler2026,
author = {Raschpichler, Matthias and Loria, Johannes Rotta Detto and Patel, Vivek and Noack, Thilo and Nüsser, Karoline and Dumpies, Oliver and Patel, Dhairya and Nagasaka, Takashi and Madaan, Prateek and Gupta, Aakriti and Thiele, Holger and Borger, Michael A and Makkar, Raj and Abdel-Wahab, Mohamed},
title = {Commissural Misalignment Following Valve‐in‐Valve Transcatheter Aortic Valve Implantation},
journal = {Catheterization and Cardiovascular Interventions},
year = {2026},
volume = {107},
number = {4},
pages = {1215--1224},
doi = {10.1002/ccd.70460}
}