Valve Performance Between Latest-Generation Balloon-Expandable and Self-Expandable Transcatheter Heart Valves in a Small Aortic Annulus
JACC: Cardiovascular Interventions · 2024
DOI: 10.1016/j.jcin.2024.08.049 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift JACC: Cardiovascular Interventions, 2024. Verfasst von Hirofumi Hioki und 24 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, IMS Tokyo Katsushika General Hospital, Tokyo, Japan. Electronic address: hioki.teikyo@gmail.com. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Prosthesis Design, Aortic Valve Stenosis, Aortic Valve. Die Arbeit wurde bisher 34-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Transcatheter aortic valve replacement (TAVR) using a self-expandable valve (SEV) promotes better hemodynamics compared with a balloon-expandable valve (BEV) in a small aortic annulus (SAA). OBJECTIVES: The authors sought to compare hemodynamic properties and clinical outcomes between the latest-generation BEV and SEV after TAVR for SAA. METHODS: We retrospectively analyzed 1,227 patients undergoing TAVR for aortic stenosis with SAA, defined as an annulus area ≤430 mm2, using the BEV (SAPIEN3 Ultra RESILIA, Edwards Lifesciences) and SEV (Evolut FX, Medtronic). The impact of valve design on severe prosthesis-patient mismatch, aortic valve mean pressure gradient ≥20 mm Hg, paravalvular leakage (PVL) ≥ mild, new permanent pacemaker implantation (PMI), and modified VARC-3 device success at discharge was evaluated using logistic regression and propensity score analysis. RESULTS: Of 1,227 patients, 798 (65.0%) underwent TAVR with BEV implantation. TAVR using BEV had a relatively higher rate of severe prosthesis-patient mismatch (OR: 1.74; 95% CI: 0.54-5.62) and significantly higher incidence of mean pressure gradient ≥20 mm Hg (OR: 2.05; 95% CI: 0.91-4.62) than that using SEV. By contrast, the BEV showed significantly lower incidence of PVL ≥ mild (OR: 0.19; 95% CI: 0.14-0.26), and new PMI (OR: 0.53; 95% CI: 0.33-0.86). The rate of device success was comparable between the BEV and the SEV. These results were confirmed by propensity score analysis. CONCLUSIONS: In TAVR for SAA, SEV demonstrated better hemodynamics than the latest BEV, whereas the latest BEV had lower incidences of PVL ≥ mild and new PMI than the SEV.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, IMS Tokyo Katsushika General Hospital, Tokyo, Japan. Electronic address: hioki.teikyo@gmail.com.
- Department of Cardiology, Toyohashi Heart Center, Toyohashi, Japan; Department of Cardiology, Nagoya Heart Center, Nagoya, Japan; Department of Cardiology, Gifu Heart Center, Gifu, Japan. Electronic address: masa-nori@n…
- Department of Cardiology, Kokura Memorial Hospital, Kokura, Japan.
- Department of Cardiology, Tokai University School of Medicine, Isehara, Japan.
- Department of Cardiology, Saiseikai Utsunomiya Hospital, Tochigi, Japan.
- Department of Cardiology, New Tokyo Hospital, Matsudo, Japan.
Zitieren
APA
Hioki, H., Yamamoto, M., Shirai, S., Ohno, Y., Yashima, F., Naganuma, T., Yamawaki, M., Watanabe, Y., Yamanaka, F., Mizutani, K., Ryuzaki, T., Noguchi, M., Izumo, M., Takagi, K., Asami, M., Ueno, H., Nishina, H., Otsuka, T., Suzuyama, H., …, & OCEAN-TAVI Investigators (2024). Valve Performance Between Latest-Generation Balloon-Expandable and Self-Expandable Transcatheter Heart Valves in a Small Aortic Annulus. JACC: Cardiovascular Interventions. https://doi.org/10.1016/j.jcin.2024.08.049
Vancouver
Hioki H, Yamamoto M, Shirai S, Ohno Y, Yashima F, Naganuma T, et al. Valve Performance Between Latest-Generation Balloon-Expandable and Self-Expandable Transcatheter Heart Valves in a Small Aortic Annulus. JACC: Cardiovascular Interventions. 2024. doi:10.1016/j.jcin.2024.08.049
BibTeX
@article{Hioki2024,
author = {Hioki, Hirofumi and Yamamoto, Masanori and Shirai, Shinichi and Ohno, Yohei and Yashima, Fumiaki and Naganuma, Toru and Yamawaki, Masahiro and Watanabe, Yusuke and Yamanaka, Futoshi and Mizutani, Kazuki and Ryuzaki, Toshinobu and Noguchi, Masahiko and Izumo, Masaki and Takagi, Kensuke and Asami, Masahiko and Ueno, Hiroshi and Nishina, Hidetaka and Otsuka, Toshiaki and Suzuyama, Hiroto and Yamasaki, Kazumasa and Nishioka, Kenji and Hachinohe, Daisuke and Fuku, Yasushi and Hayashida, Kentaro and OCEAN-TAVI Investigators, OCEAN-TAVI Investigators},
title = {Valve Performance Between Latest-Generation Balloon-Expandable and Self-Expandable Transcatheter Heart Valves in a Small Aortic Annulus},
journal = {JACC: Cardiovascular Interventions},
year = {2024},
doi = {10.1016/j.jcin.2024.08.049}
}
EUVIMED