Long‐term durability of self‐expanding and balloon‐expandable transcatheter aortic valve prostheses: UK TAVI registry
Catheterization and Cardiovascular Interventions · 2023
DOI: 10.1002/ccd.30627 ↗At a glance
Journal article in Catheterization and Cardiovascular Interventions, 2023. Written by Noman Ali and 17 co-authors. Institution: Department of Cardiology, Leeds Teaching Hospitals NHS Trust, Leeds, UK. Major MEDLINE topics: Transcatheter Aortic Valve Replacement, Aortic Valve Stenosis, Heart Valve Prosthesis. Cited 25 times so far (Crossref); 24 references.
Abstract (original language)
Background
With expansion of transcatheter aortic valve implantation (TAVI) into younger patients, valve durability is critically important.
Aims
We aimed to evaluate long‐term valve function and incidence of severe structural valve deterioration (SVD) among patients ≥ 10‐years post‐TAVI and with echocardiographic follow‐up at least 5‐years postprocedure.
Methods
Data on patients who underwent TAVI from 2007 to 2011 were obtained from the UK TAVI registry. Patients with paired echocardiograms postprocedure and ≥5‐years post‐TAVI were included. Severe SVD was determined according to European task force guidelines.
Results
221 patients (79.4 ± 7.3 years; 53% male) were included with median echocardiographic follow‐up 7.0 years (range 5−13 years). Follow‐up exceeded 10 years in 43 patients (19.5%). Valve types were the supra‐annular self‐expanding CoreValve (SEV; n = 143, 67%), balloon‐expandable SAPIEN/XT (BEV; n = 67, 31%), Portico ( n = 4, 5%) and unknown ( n = 7, 3%). There was no difference between postprocedure and follow‐up peak gradient in the overall cohort (19.3 vs. 18.4 mmHg; p = NS) or in those with ≥10‐years follow‐up (21.1 vs. 21.1 mmHg; p = NS). Severe SVD occurred in 13 patients (5.9%; median 7.8‐years post‐TAVI). Three cases (23.1%) were due to regurgitation and 10 (76.9%) to stenosis. Valve‐related reintervention/death occurred in 5 patients (2.3%). Severe SVD was more frequent with BEV than SEV (11.9% vs. 3.5%; p = 0.02), driven by a difference in patients treated with small valves (BEV 28.6% vs. SEV 3.0%; p < 0.01).
Conclusions
Hemodynamic function of transcatheter heart valves remains stable up to more than 10 years post‐TAVI. Severe SVD occurred in 5.9%, and valve‐related death/reintervention in 2.3%. Severe SVD was more common with BEV than SEV.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiology, Leeds Teaching Hospitals NHS Trust, Leeds, UK.
- Department of Cardiology, Royal Sussex County Hospital, Brighton, UK.
- Department of Cardiology, Royal Victoria Hospital, Belfast, UK.
- Department of Cardiology, King's College Hospital, London, UK.
- Department of Cardiology, St. George's Hospital, London, UK.
- Department of Cardiology, Royal Papworth Hospital NHS Foundation Trust, Cambridge, UK.
ORCID profiles: Noman Ali, Jonathan Mailey, Andrew Wiper, Daniel J Blackman
Cite
APA
Ali, N., Hildick-Smith, D., Parker, J., Malkin, C. J., Cunnington, M. S., Gurung, S., Mailey, J., MacCarthy, P. A., Bharucha, A., Brecker, S. J., Hoole, S. P., Dorman, S., Doshi, S. N., Wiper, A., Buch, M. H., Banning, A. P., Spence, M. S., & Blackman, D. J. (2023). Long‐term durability of self‐expanding and balloon‐expandable transcatheter aortic valve prostheses: UK TAVI registry. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.30627
Vancouver
Ali N, Hildick-Smith D, Parker J, Malkin CJ, Cunnington MS, Gurung S, et al. Long‐term durability of self‐expanding and balloon‐expandable transcatheter aortic valve prostheses: UK TAVI registry. Catheterization and Cardiovascular Interventions. 2023. doi:10.1002/ccd.30627
BibTeX
@article{Ali2023,
author = {Ali, Noman and Hildick-Smith, David and Parker, Jessica and Malkin, Christopher J and Cunnington, Michael S and Gurung, Shuslim and Mailey, Jonathan and MacCarthy, Philip A and Bharucha, Apurva and Brecker, Stephen J and Hoole, Stephen P and Dorman, Stephen and Doshi, Sagar N and Wiper, Andrew and Buch, Mamta H and Banning, Adrian P and Spence, Mark S and Blackman, Daniel J},
title = {Long‐term durability of self‐expanding and balloon‐expandable transcatheter aortic valve prostheses: UK TAVI registry},
journal = {Catheterization and Cardiovascular Interventions},
year = {2023},
doi = {10.1002/ccd.30627}
}
EUVIMED