EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

A Wearable Touch‐Activated Device Integrated with Hollow Microneedles for Continuous Sampling and Sensing of Dermal Interstitial Fluid

Taher Abbasiasl iD, Fariborz Mirlou iD, Hadi Mirzajani iD, Mohammad Javad Bathaei iD, Emin Istif iD, Narges Shomalizadeh iD, Rümeysa Emine Cebecioğlu iD, Ecem Ezgi Özkahraman, Umut Can Yener iD, Levent Beker iD

Advanced Materials · 2024-01 · 36 (2)

DOI: 10.1002/adma.202304704 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advanced Materials, online erschienen am 23.11.2023. Verfasst von Taher Abbasiasl und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Sciences and Engineering, Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey. Hauptthemen laut MEDLINE: Extracellular Fluid, Wearable Electronic Devices. Die Arbeit wurde bisher 89-mal zitiert (Crossref) und nennt 62 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: The Scientific and Technological Research Council of Turkey, HORIZON EUROPE European Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Dermal interstitial fluid (ISF) is emerging as a rich source of biomarkers that complements conventional biofluids such as blood and urine. However, the impact of ISF sampling in clinical applications has been limited owing to the challenges associated with extraction. The implementation of microneedle‐based wearable devices that can extract dermal ISF in a pain‐free and easy‐to‐use manner has attracted growing attention in recent years. Here, a fully integrated touch‐activated wearable device based on a laser‐drilled hollow microneedle (HMN) patch for continuous sampling and sensing of dermal ISF is introduced. The developed platform can produce and maintain the required vacuum pressure (as low as ≈ −53 kPa) to collect adequate volumes of ISF (≈2 µL needle −1 h −1 ) for medical applications. The vacuum system can be activated through a one‐touch finger operation. A parametric study is performed to investigate the effect of microneedle array size, vacuum pressure, and extraction duration on collected ISF. The capability of the proposed platform for continuous health monitoring is further demonstrated by the electrochemical detection of glucose and pH levels of ISF in animal models. This HMN‐based system provides an alternative tool to the existing invasive techniques for ISF collection and sensing for medical diagnosis and treatment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Extracellular FluidWearable Electronic DevicesAnimalsTouchGlucoseNeedles

Stichwörter der Autor:innen

ISF samplingcontinuous monitoringelectrochemical sensinghollow microneedlesinterstitial fluid

Substanzen

Glucose

Autor:innen und Institutionen

  • Department of Biomedical Sciences and Engineering, Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey.
  • Department of Mechanical Engineering, Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey.
  • Faculty of Engineering and Natural Sciences, Kadir Has University, Cibali, Istanbul, 34083, Turkey.
  • Koç University Research Center for Translational Research (KUTTAM), Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey.
  • Department of Material Science and Engineering, Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey.
  • Nanofabrication and Nanocharacterization Center for Scientific and Technological Advanced Research (n2Star), Koç University, Rumelifeneri Yolu, Sarıyer, Istanbul, 34450, Turkey.

ORCID-Profile: Taher Abbasiasl, Fariborz Mirlou, Hadi Mirzajani, Mohammad Javad Bathaei, Emin Istif, Narges Shomalizadeh, Rümeysa Emine Cebecioğlu, Umut Can Yener, Levent Beker

Förderung

  • The Scientific and Technological Research Council of Turkey
  • HORIZON EUROPE European Research Council

Zitieren

APA

Abbasiasl, T., Mirlou, F., Mirzajani, H., Bathaei, M. J., Istif, E., Shomalizadeh, N., Cebecioğlu, R. E., Özkahraman, E. E., Yener, U. C., & Beker, L. (2024). A Wearable Touch‐Activated Device Integrated with Hollow Microneedles for Continuous Sampling and Sensing of Dermal Interstitial Fluid. Advanced Materials, 36(2). https://doi.org/10.1002/adma.202304704

Vancouver

Abbasiasl T, Mirlou F, Mirzajani H, Bathaei MJ, Istif E, Shomalizadeh N, et al. A Wearable Touch‐Activated Device Integrated with Hollow Microneedles for Continuous Sampling and Sensing of Dermal Interstitial Fluid. Advanced Materials. 2024;36(2). doi:10.1002/adma.202304704

BibTeX
@article{Abbasiasl2024,
  author = {Abbasiasl, Taher and Mirlou, Fariborz and Mirzajani, Hadi and Bathaei, Mohammad Javad and Istif, Emin and Shomalizadeh, Narges and Cebecioğlu, Rümeysa Emine and Özkahraman, Ecem Ezgi and Yener, Umut Can and Beker, Levent},
  title = {A Wearable Touch‐Activated Device Integrated with Hollow Microneedles for Continuous Sampling and Sensing of Dermal Interstitial Fluid},
  journal = {Advanced Materials},
  year = {2024},
  volume = {36},
  number = {2},
  doi = {10.1002/adma.202304704}
}
Kostenlos starten →