Assessment of image quality and detectability in computed tomography: Inter‐laboratory comparison of Fourier‐based metrics

Pascal Monnin iD, Damien Racine, Luis Ammour, Yves Barbotteau, Davide Cester iD, Dimitar Petrov, Kwinten Torfs iD, Raffaele Villa iD, Nicoletta Paruccini, Joël Greffier iD, Anaïs Viry

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70838 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Pascal Monnin und 10 weiteren Autor:innen. Beteiligte Einrichtung: Institute of radiation physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Fourier Analysis, Phantoms, Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The task‐based assessment of image quality in computed tomography (CT) is increasingly used for benchmarking, optimizing acquisition and reconstruction parameters, and predicting observers’ performance. To achieve consistent and reproducible results rely on standardized methodologies and guidelines.

Purpose

This study performed an inter‐laboratory comparison to evaluate the consistency of CT image quality metrics and assess inter‐laboratory variability.

Methods

Six series of images of a Catphan 500 phantom were acquired at three dose levels (1.5/5.0/15.0 mGy) with two (standard and high‐resolution) kernels and sent to seven independent laboratories. Participants calculated the task transfer function (TTF) for spatial resolution and noise power spectrum (NPS) for noise magnitude and texture using standardized Fourier‐based analysis on the shared images. The TTFs and NPS were used in the non‐prewhitening observer model without (NPW) and with eye filter (NPWE) to calculate a detectability index for small (2 mm) and large targets (10 mm).

Results

The difference observed in TTFs resulted mainly from the different noise reduction techniques used during the calculation. The frequency corresponding to the TTF value 50% varied from 0.27 to 0.43 mm −1 for the standard images and from 0.73 to 0.92 mm −1 for the high‐resolution images. Variability in NPS primarily stemmed from the choice of regions of interest location in non‐stationary noise images that lead to magnitude differences up to 40% for the standard kernel and 70% for the high‐resolution kernel. The combined differences in TTF and NPS resulted in differences up to 39% in the NPW detectability index. The large variability in the human visual transfer function and visualization parameters increased the differences by a factor of 6.25 in the NPWE model and highlighted the lack of standardization.

Conclusion

Adapting TTF and NPS to a more complex and localized analysis of CT images with non‐uniform CNRs, rather than a simple global evaluation, increases computational variability and requires a move towards more standardization of metrics. In the absence of complete standardization of image quality measurement, documentation of methods and all calculation parameters is essential for more robust evaluation of CT image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedFourier AnalysisPhantoms, ImagingImage Processing, Computer-AssistedAlgorithmsRadiographic Image Interpretation, Computer-AssistedHumansObserver Variation

Stichwörter der Autor:innen

Fourier‐based metricscomputed tomographydetectabilityimage qualityinter‐laboratory comparison

Autor:innen und Institutionen

  • Institute of radiation physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne, Lausanne, Switzerland.
  • CHU Nantes, Nantes Université, Nantes, France.
  • Service d'imagerie-Hôpital Clairval, Ramsay Santé, Marseille, France.
  • Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zürich, Switzerland.
  • Department of Imaging and Pathology, KU Leuven, Leuven, Belgium.
  • Medical Physics Department, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.

ORCID-Profile: Pascal Monnin, Davide Cester, Kwinten Torfs, Raffaele Villa, Joël Greffier

Zitieren

APA

Monnin, P., Racine, D., Ammour, L., Barbotteau, Y., Cester, D., Petrov, D., Torfs, K., Villa, R., Paruccini, N., Greffier, J., & Viry, A. (2026). Assessment of image quality and detectability in computed tomography: Inter‐laboratory comparison of Fourier‐based metrics. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70838

Vancouver

Monnin P, Racine D, Ammour L, Barbotteau Y, Cester D, Petrov D, et al. Assessment of image quality and detectability in computed tomography: Inter‐laboratory comparison of Fourier‐based metrics. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70838

BibTeX
@article{Monnin2026,
  author = {Monnin, Pascal and Racine, Damien and Ammour, Luis and Barbotteau, Yves and Cester, Davide and Petrov, Dimitar and Torfs, Kwinten and Villa, Raffaele and Paruccini, Nicoletta and Greffier, Joël and Viry, Anaïs},
  title = {Assessment of image quality and detectability in computed tomography: Inter‐laboratory comparison of Fourier‐based metrics},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70838}
}
Kostenlos starten →