Assessing the role of plan complexity and target geometry through multi‐institutional gel based end‐to‐end QA in multi‐focal single isocenter stereotactic radiosurgery

Despoina Stasinou iD, Kalliopi Platoni iD, Vasiliki Margaroni iD, Kyveli Zourari, Emmanouil Zoros, Niko Papanikolaou, Evangelos Pappas

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70758 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Despoina Stasinou und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Sciences, Radiology and Radiotherapy Sector, University of West Attica, Athens, Attica, Greece. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy Planning, Computer-Assisted, Brain Neoplasms, Quality Assurance, Health Care. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Single‐isocenter stereotactic radiosurgery enables efficient treatment of multiple brain metastases (SI‐MBM SRS), but demands high geometric accuracy. Plan complexity metrics are increasingly used as indicators for quality assurance (QA) performance, however their applicability to SRS remains uncertain, particularly in the context of multi‐institutional variability.

Purpose

This study evaluated the relationship between plan complexity, target geometry, and end‐to‐end dosimetric QA outcomes for SI‐MBM SRS across multiple institutions.

Methods

Forty‐two SI‐MBM SRS plans from different centers and platforms were delivered to polymer gel phantoms, providing high‐resolution 3D dose measurements. Gamma passing rates (GPRs) were calculated under 3%/2 mm, 5%/2 mm and 5%/1 mm criteria and were correlated with eleven established complexity metrics calculated per plan. Geometric factors, including target equivalent diameter and distance‐to‐isocenter, were analyzed. Receiver‐operating‐characteristic (ROC) analysis was performed to identify optimal thresholds for predicting QA pass/fail (≥90% GPR).

Results

No statistically significant differences between the different linacs and treatment planning systems were found, nor strong or consistent correlations between complexity metrics and GPRs. In contrast, geometric parameters were more influential: off‐axis distance and target size significantly affected QA performance, with the largest differences observed for far‐off‐axis lesions. ROC analysis identified optimal thresholds of 34.9 mm for distance‐to‐isocenter and 4.8 mm for equivalent diameter in predicting QA outcomes (AUC∼0.60‐0.65), although the predictive performance remained modest.

Conclusions

In this multi‐institutional, gel‐based end‐to‐end study of SI‐MBM SRS, target geometry was a stronger QA performance predictor than complexity metrics. These findings emphasize the importance of geometry‐aware QA strategies and the need for further standardized, multi‐institutional evaluations to clarify the interplay between complexity, geometry, and machine performance in SRS.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy Planning, Computer-AssistedBrain NeoplasmsQuality Assurance, Health CarePhantoms, ImagingHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedParticle AcceleratorsOrgans at RiskGels

Stichwörter der Autor:innen

anthropomorphic phantomcomplexity metricsend‐to‐end testgel dosimetrymultiple brain metastasesmulti‐institutional studysingle isocenter SRS

Substanzen

Gels

Autor:innen und Institutionen

  • Department of Biomedical Sciences, Radiology and Radiotherapy Sector, University of West Attica, Athens, Attica, Greece.
  • Department of Applied Medical Physics, Medical School, Attikon University Hospital, National and Kapodistrian University of Athens, Athens, Attica, Greece.
  • Medical Physics Laboratory, School of Medicine, National and Kapodistrian University of Athens, Athens, Attica, Greece.
  • Department of Applied Medical Physics, University General Hospital of West Attica "H AGIA VARBARA", Athens, Attica, Greece.
  • Department of Radiation Oncology, The University of Texas Health Science Center at San Antonio, San Antonio, Texas, USA.
  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, England, UK.

ORCID-Profile: Despoina Stasinou, Kalliopi Platoni, Vasiliki Margaroni

Zitieren

APA

Stasinou, D., Platoni, K., Margaroni, V., Zourari, K., Zoros, E., Papanikolaou, N., & Pappas, E. (2026). Assessing the role of plan complexity and target geometry through multi‐institutional gel based end‐to‐end QA in multi‐focal single isocenter stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70758

Vancouver

Stasinou D, Platoni K, Margaroni V, Zourari K, Zoros E, Papanikolaou N, et al. Assessing the role of plan complexity and target geometry through multi‐institutional gel based end‐to‐end QA in multi‐focal single isocenter stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70758

BibTeX
@article{Stasinou2026,
  author = {Stasinou, Despoina and Platoni, Kalliopi and Margaroni, Vasiliki and Zourari, Kyveli and Zoros, Emmanouil and Papanikolaou, Niko and Pappas, Evangelos},
  title = {Assessing the role of plan complexity and target geometry through multi‐institutional gel based end‐to‐end QA in multi‐focal single isocenter stereotactic radiosurgery},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70758}
}
Kostenlos starten →