Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70239 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Minsoo Chun und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea. Hauptthemen laut MEDLINE: Phantoms, Imaging, Cone-Beam Computed Tomography, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Korea government (MSIT), Disabled Enterprise Business Center.
Abstract (Originalsprache)
Purpose
This study presents MZ CUBE, a newly developed multipurpose phantom designed to assess both image quality and geometrical accuracy in image‐guided radiotherapy.
Methods
The phantom is a 16 × 16 × 16 cm 3 cube made of acrylonitrile butadiene styrene and designed to evaluate HU constancy, HU uniformity, spatial integrity, spatial resolution, low contrast resolution, and geometrical accuracy. Two CT simulators (Big Bore RT and Discovery RT Gen3) and two cone‐beam CT (CBCT) systems (XVI and OBI) were used for verification alongside commercial phantoms (ACR 464 and CatPhan 504).
Results
HU constancy differences were within 3 HU, and uniformity was within 1 HU for CT and 2 HU for CBCT, based on repeated MZ CUBE measurements. Spatial resolution with the MZ CUBE was 7 lp/cm (Big Bore RT), 8 lp/cm (Discovery RT Gen3), 7 lp/cm (XVI), and 8 lp/cm (OBI). The minimum discernible disk diameters were 7, 5, 6, and 6 mm, respectively. Geometrical accuracy was confirmed with translational offsets below 1 mm and rotational deviations below 0.2° for both CBCT systems.
Conclusions
MZ CUBE enables comprehensive image quality and geometric accuracy evaluation in a single scan, offering an efficient and cost‐effective solution for radiotherapy QA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea.
- Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
- Department of Radiation Oncology, Chung-Ang University College of Medicine, Seoul, Republic of Korea.
- Paprica Lab Co., Ltd, Seoul, Republic of Korea.
- Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea.
- Department of Radiation Oncology, Seoul National University College of Medicine, Seoul, Republic of Korea.
ORCID-Profile: Minsoo Chun, Jaewon Hong, Jin Hwa Choi, Chang-Heon Choi, Jung-In Kim, Hong-Gyun Wu, Minjae Choi, Na Young Jang, So-Yeon Park
Förderung
- National Research Foundation of Korea
- Korea government (MSIT)
- Disabled Enterprise Business Center
- Ministry of SMEs and Startups
Zitieren
APA
Chun, M., Hong, J., Choi, J. H., Choi, C. H., Kim, J. I., Wu, H. G., Choi, M., Jang, N. Y., & Park, S. Y. (2025). Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70239
Vancouver
Chun M, Hong J, Choi JH, Choi CH, Kim JI, Wu HG, et al. Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70239
BibTeX
@article{Chun2025,
author = {Chun, Minsoo and Hong, Jaewon and Choi, Jin Hwa and Choi, Chang-Heon and Kim, Jung-In and Wu, Hong-Gyun and Choi, Minjae and Jang, Na Young and Park, So-Yeon},
title = {Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70239}
}