EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy

Minsoo Chun iD, Jaewon Hong iD, Jin Hwa Choi iD, Chang-Heon Choi iD, Jung-In Kim iD, Hong-Gyun Wu iD, Minjae Choi iD, Na Young Jang iD, So-Yeon Park iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70239 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Minsoo Chun und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea. Hauptthemen laut MEDLINE: Phantoms, Imaging, Cone-Beam Computed Tomography, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Korea government (MSIT), Disabled Enterprise Business Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study presents MZ CUBE, a newly developed multipurpose phantom designed to assess both image quality and geometrical accuracy in image‐guided radiotherapy.

Methods

The phantom is a 16 × 16 × 16 cm 3 cube made of acrylonitrile butadiene styrene and designed to evaluate HU constancy, HU uniformity, spatial integrity, spatial resolution, low contrast resolution, and geometrical accuracy. Two CT simulators (Big Bore RT and Discovery RT Gen3) and two cone‐beam CT (CBCT) systems (XVI and OBI) were used for verification alongside commercial phantoms (ACR 464 and CatPhan 504).

Results

HU constancy differences were within 3 HU, and uniformity was within 1 HU for CT and 2 HU for CBCT, based on repeated MZ CUBE measurements. Spatial resolution with the MZ CUBE was 7 lp/cm (Big Bore RT), 8 lp/cm (Discovery RT Gen3), 7 lp/cm (XVI), and 8 lp/cm (OBI). The minimum discernible disk diameters were 7, 5, 6, and 6 mm, respectively. Geometrical accuracy was confirmed with translational offsets below 1 mm and rotational deviations below 0.2° for both CBCT systems.

Conclusions

MZ CUBE enables comprehensive image quality and geometric accuracy evaluation in a single scan, offering an efficient and cost‐effective solution for radiotherapy QA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingCone-Beam Computed TomographyRadiotherapy Planning, Computer-AssistedQuality Assurance, Health CareImage Processing, Computer-AssistedRadiotherapy, Image-GuidedTomography, X-Ray ComputedNeoplasmsHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

geometrical accuracyimage qualityimage‐guided radiation therapyquality assurance phantom

Autor:innen und Institutionen

  • Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea.
  • Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
  • Department of Radiation Oncology, Chung-Ang University College of Medicine, Seoul, Republic of Korea.
  • Paprica Lab Co., Ltd, Seoul, Republic of Korea.
  • Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea.
  • Department of Radiation Oncology, Seoul National University College of Medicine, Seoul, Republic of Korea.

ORCID-Profile: Minsoo Chun, Jaewon Hong, Jin Hwa Choi, Chang-Heon Choi, Jung-In Kim, Hong-Gyun Wu, Minjae Choi, Na Young Jang, So-Yeon Park

Förderung

  • National Research Foundation of Korea
  • Korea government (MSIT)
  • Disabled Enterprise Business Center
  • Ministry of SMEs and Startups

Zitieren

APA

Chun, M., Hong, J., Choi, J. H., Choi, C. H., Kim, J. I., Wu, H. G., Choi, M., Jang, N. Y., & Park, S. Y. (2025). Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70239

Vancouver

Chun M, Hong J, Choi JH, Choi CH, Kim JI, Wu HG, et al. Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70239

BibTeX
@article{Chun2025,
  author = {Chun, Minsoo and Hong, Jaewon and Choi, Jin Hwa and Choi, Chang-Heon and Kim, Jung-In and Wu, Hong-Gyun and Choi, Minjae and Jang, Na Young and Park, So-Yeon},
  title = {Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT for radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70239}
}
Kostenlos starten →