EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

The role of abdominal compression in the reduction of respiratory motion for cardiac radioablation

Jakob Marshall iD, Alanah Bergman iD, Tania Karan iD, Marc W Deyell, Devin Schellenberg, Steven Thomas

Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)

DOI: 10.1002/acm2.70561 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.03.2026. Verfasst von Jakob Marshall und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University of British Columbia, Vancouver, British Columbia, Canada. Hauptthemen laut MEDLINE: Respiration, Abdomen, Radiosurgery, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Vancouver Coastal Health Research Institute, Providence Health Care Research Institute, BC Cancer Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Research on cardiac radioablation (CR) has highlighted the importance of accounting for respiratory‐induced motion during treatment, promoting investigation into methods to reduce its impact.

Purpose

To investigate the efficacy of abdominal compression (AC) to reduce respiratory‐induced motion for CR patients by quantifying their respiratory‐induced motion under AC and free breathing (FB) conditions. Methods and materials Nine patients were imaged with 5 Hz bi‐planar fluoroscopy on the Vero4DRT linear accelerator for 15–20 seconds under both AC and FB conditions in preparation for CR. As the soft tissue target is not visible on bi‐planar X‐ray images, the implantable cardioverter defibrillator (ICD) lead tip was used as a motion surrogate. The position of the ICD lead tip was tracked and triangulated in each bi‐planar image frame, with the respiratory component of motion extracted using a lowpass filter. Properties of respiratory‐induced motion were quantified including the average (across patient breaths) peak‐to‐peak motion extent, the mean total 3D magnitude of displacement, and the standard deviation of motion for each patient.

Results

The average (95% CI) extent of RV lead respiratory‐induced motion was [3.0 (1.1, 5.0), 2.8 (1.6, 4.1), 7.0 (4.2, 9.7)] mm under FB and [2.1 (1.4, 2.8), 2.7 (1.7, 3.8), 4.7 (3.5, 5.9)] mm under AC in the [right–left, anterior–posterior, inferior–superior] directions. Eight out of nine patients had a reduction in the mean respiratory‐induced total 3D displacement under AC, with an average ( STD) across patients of (3.2 1.7) mm under FB and (2.2 0.8) mm under AC. Averaged across patients, the standard deviation (95% CI) of respiratory‐induced motion was [1.3 (0.5, 2.0), 1.2 (0.7, 1.6), 3.1 (1.7, 4.5)] mm under FB and [0.9 (0.6, 1.3), 1.2 (0.6, 1.8), 2.0 (1.3, 2.7)] mm under AC.

Conclusions

For eight out of nine CR patients investigated, the respiratory‐induced RV lead motion was reduced when using AC compared to FB conditions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RespirationAbdomenRadiosurgeryRadiotherapy Planning, Computer-AssistedHumansFluoroscopyMovementRadiotherapy DosageDefibrillators, ImplantableImage Processing, Computer-Assisted

Stichwörter der Autor:innen

abdominal compressionrespiratory motionstereotactic arrhythmia radioablation

Autor:innen und Institutionen

  • Department of Physics, University of British Columbia, Vancouver, British Columbia, Canada.
  • Medical Physics, BC Cancer, Vancouver, British Columbia, Canada.
  • Department of Surgery, Division of Radiation Oncology and Experimental Radiotherapeutics, University of British Columbia, Vancouver, British Columbia, Canada.
  • Centre for Cardiovascular Innovation and Division of Cardiology, University of British Columbia, Vancouver, British Columbia, Canada.
  • Radiation Oncology, BC Cancer, Surrey, British Columbia, Canada.

ORCID-Profile: Jakob Marshall, Alanah Bergman, Tania Karan

Förderung

  • Vancouver Coastal Health Research Institute
  • Providence Health Care Research Institute
  • BC Cancer Foundation

Zitieren

APA

Marshall, J., Bergman, A., Karan, T., Deyell, M. W., Schellenberg, D., & Thomas, S. (2026). The role of abdominal compression in the reduction of respiratory motion for cardiac radioablation. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70561

Vancouver

Marshall J, Bergman A, Karan T, Deyell MW, Schellenberg D, Thomas S. The role of abdominal compression in the reduction of respiratory motion for cardiac radioablation. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70561

BibTeX
@article{Marshall2026,
  author = {Marshall, Jakob and Bergman, Alanah and Karan, Tania and Deyell, Marc W and Schellenberg, Devin and Thomas, Steven},
  title = {The role of abdominal compression in the reduction of respiratory motion for cardiac radioablation},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {4},
  doi = {10.1002/acm2.70561}
}
Kostenlos starten →