To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging

Vanessa Da Silva Mendes iD, Sylvia Garny, Lili Huang, Stephan Schönecker, Christian Trapp, Frederik Fuchs, Christopher Kurz, Claus Belka, Guillaume Landry iD, Michael Reiner, Stefanie Corradini

Journal of Applied Clinical Medical Physics · 2026-03 · 27 (3)

DOI: 10.1002/acm2.70505 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 08.03.2026. Verfasst von Vanessa Da Silva Mendes und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany. Hauptthemen laut MEDLINE: Radiosurgery, Brain Neoplasms, Phantoms, Imaging, Patient Positioning. Die Arbeit nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Brainlab SE.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Frameless linear accelerator (linac)‐based image‐guided stereotactic radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are a widely used treatment option for intracranial lesions. Given the high radiation doses involved, it is crucial to maintain precise patient positioning throughout treatment. This requires that geometric inaccuracies arising from patient motion or setup errors are identified and corrected. With frameless immobilization, image‐guidance has a greater impact, especially in non‐coplanar settings that can lead to patient motion and discrepancies between couch and radiation isocenters.

Purpose

Both patient and phantom studies were conducted to assess and quantify the magnitude of geometric uncertainties after couch rotations, aiming at evaluating the clinical need for their correction to warrant a precise treatment delivery.

Methods

Intrafraction X‐ray data, performed by ExacTrac Dynamic (ETD) to monitor and correct patients’ position throughout treatment delivery, were collected from 50 patients treated for brain metastases in stereotactic non‐coplanar schemes and immobilized by stereotactic double‐layered thermoplastic mask systems: 26 patients treated in 40 single‐fraction SRS (168 stereoscopic X‐ray images); 24 treated with FSRT in 128 fractions (278 stereoscopic X‐ray images). Additionally, a head phantom was utilized and 350 measurements under two different couch loads were carried out to distinguish true patient motion from deviations caused by couch rotations or system‐related effects. For both studies, ETD stereoscopic X‐rays were acquired after each couch rotation and the first measured positioning deviation was calculated by comparing X‐ray images to the treatment plan's digitally reconstructed radiographs.

Results

Clinically relevant deviations were observed, exceeding clinical tolerance (≥ 0.5 mm/0.5°) mostly in the lateral and yaw directions and requiring repositioning in nearly half of the couch rotations. These deviations measuring up to 2 mm, revealed to be emerging mainly from patient motion rather than linac setup, as the phantom study showed maximum deviations of up to 0.6 mm and 0.4° when simulating a patient treatment and an interquartile range that did not exceed 0.2 mm and 0.2°.

Conclusions

These findings demonstrate the importance of a continuous intrafraction motion monitoring and repositioning in cranial stereotactic treatments, especially in non‐coplanar settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryBrain NeoplasmsPhantoms, ImagingPatient PositioningRadiotherapy Planning, Computer-AssistedRadiotherapy Setup ErrorsTomography, X-Ray ComputedRadiotherapy, Image-GuidedHumansParticle AcceleratorsRadiotherapy, Intensity-ModulatedRadiotherapy DosageRotationImmobilizationMovement

Stichwörter der Autor:innen

IGRTSRSX‐ray intrafraction monitoringintracranialnon‐coplanar beamspatient (re‐)positioningstereoscopic X‐ray imaging

Autor:innen und Institutionen

  • Department of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
  • German Cancer Consortium (DKTK), partner site Munich, a partnership between DKFZ and LMU University Hospital, Munich, Bavaria, Germany.
  • Bavarian Cancer Research Centre (BZKF), Munich, Bavaria, Germany.

ORCID-Profile: Vanessa Da Silva Mendes, Guillaume Landry

Förderung

  • Brainlab SE

Zitieren

APA

Da Silva Mendes, V., Garny, S., Huang, L., Schönecker, S., Trapp, C., Fuchs, F., Kurz, C., Belka, C., Landry, G., Reiner, M., & Corradini, S. (2026). To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging. Journal of Applied Clinical Medical Physics, 27(3). https://doi.org/10.1002/acm2.70505

Vancouver

Da Silva Mendes V, Garny S, Huang L, Schönecker S, Trapp C, Fuchs F, et al. To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging. Journal of Applied Clinical Medical Physics. 2026;27(3). doi:10.1002/acm2.70505

BibTeX
@article{DaSilvaMendes2026,
  author = {Da Silva Mendes, Vanessa and Garny, Sylvia and Huang, Lili and Schönecker, Stephan and Trapp, Christian and Fuchs, Frederik and Kurz, Christopher and Belka, Claus and Landry, Guillaume and Reiner, Michael and Corradini, Stefanie},
  title = {To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {3},
  doi = {10.1002/acm2.70505}
}
Kostenlos starten →