To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging
Journal of Applied Clinical Medical Physics · 2026-03 · 27 (3)
DOI: 10.1002/acm2.70505 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 08.03.2026. Verfasst von Vanessa Da Silva Mendes und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany. Hauptthemen laut MEDLINE: Radiosurgery, Brain Neoplasms, Phantoms, Imaging, Patient Positioning. Die Arbeit nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Brainlab SE.
Abstract (Originalsprache)
Background
Frameless linear accelerator (linac)‐based image‐guided stereotactic radiosurgery (SRS) or fractionated stereotactic radiotherapy (FSRT) are a widely used treatment option for intracranial lesions. Given the high radiation doses involved, it is crucial to maintain precise patient positioning throughout treatment. This requires that geometric inaccuracies arising from patient motion or setup errors are identified and corrected. With frameless immobilization, image‐guidance has a greater impact, especially in non‐coplanar settings that can lead to patient motion and discrepancies between couch and radiation isocenters.
Purpose
Both patient and phantom studies were conducted to assess and quantify the magnitude of geometric uncertainties after couch rotations, aiming at evaluating the clinical need for their correction to warrant a precise treatment delivery.
Methods
Intrafraction X‐ray data, performed by ExacTrac Dynamic (ETD) to monitor and correct patients’ position throughout treatment delivery, were collected from 50 patients treated for brain metastases in stereotactic non‐coplanar schemes and immobilized by stereotactic double‐layered thermoplastic mask systems: 26 patients treated in 40 single‐fraction SRS (168 stereoscopic X‐ray images); 24 treated with FSRT in 128 fractions (278 stereoscopic X‐ray images). Additionally, a head phantom was utilized and 350 measurements under two different couch loads were carried out to distinguish true patient motion from deviations caused by couch rotations or system‐related effects. For both studies, ETD stereoscopic X‐rays were acquired after each couch rotation and the first measured positioning deviation was calculated by comparing X‐ray images to the treatment plan's digitally reconstructed radiographs.
Results
Clinically relevant deviations were observed, exceeding clinical tolerance (≥ 0.5 mm/0.5°) mostly in the lateral and yaw directions and requiring repositioning in nearly half of the couch rotations. These deviations measuring up to 2 mm, revealed to be emerging mainly from patient motion rather than linac setup, as the phantom study showed maximum deviations of up to 0.6 mm and 0.4° when simulating a patient treatment and an interquartile range that did not exceed 0.2 mm and 0.2°.
Conclusions
These findings demonstrate the importance of a continuous intrafraction motion monitoring and repositioning in cranial stereotactic treatments, especially in non‐coplanar settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, LMU University Hospital, Munich, Bavaria, Germany.
- German Cancer Consortium (DKTK), partner site Munich, a partnership between DKFZ and LMU University Hospital, Munich, Bavaria, Germany.
- Bavarian Cancer Research Centre (BZKF), Munich, Bavaria, Germany.
ORCID-Profile: Vanessa Da Silva Mendes, Guillaume Landry
Förderung
- Brainlab SE
Zitieren
APA
Da Silva Mendes, V., Garny, S., Huang, L., Schönecker, S., Trapp, C., Fuchs, F., Kurz, C., Belka, C., Landry, G., Reiner, M., & Corradini, S. (2026). To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging. Journal of Applied Clinical Medical Physics, 27(3). https://doi.org/10.1002/acm2.70505
Vancouver
Da Silva Mendes V, Garny S, Huang L, Schönecker S, Trapp C, Fuchs F, et al. To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging. Journal of Applied Clinical Medical Physics. 2026;27(3). doi:10.1002/acm2.70505
BibTeX
@article{DaSilvaMendes2026,
author = {Da Silva Mendes, Vanessa and Garny, Sylvia and Huang, Lili and Schönecker, Stephan and Trapp, Christian and Fuchs, Frederik and Kurz, Christopher and Belka, Claus and Landry, Guillaume and Reiner, Michael and Corradini, Stefanie},
title = {To shift or not to shift: identifying and correcting patient motion after couch rotations in non‐coplanar intracranial radiosurgery with stereoscopic X‐ray imaging},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {3},
doi = {10.1002/acm2.70505}
}
EUVIMED