Three‐dimensional surface scanning of patients to extend computed tomography field of view and scan length for dose calculation and collision avoidance in radiotherapy treatment planning
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70333 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Amanda M Jackson und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, WashU Medicine, 4921 Parkview Place, St. Louis, Missouri, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Tomography, X-Ray Computed, Imaging, Three-Dimensional, Phantoms, Imaging. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Radiotherapy treatment planning requires knowledge of a patient's anatomical surface for use in dose calculations, collision checks, or surface‐guided radiation therapy (SGRT). This information is usually obtained from a computed tomography (CT) scan which may encounter field‐of‐view (FOV) limitations or require additional radiation dose beyond the treatment area to capture the full patient habitus.
Purpose
This study demonstrates an affordable alternative solution to obtain a patient's surface information without CT or room mounted surface‐guidance imaging cameras for the purpose of detecting patient collisions and providing additional skin surface information needed for appropriate depth calculations of clipped FOV scans.
Methods
Three‐dimensional (3D) surface models of five patients were acquired immediately prior to CT simulation using an iPad Air 2 equipped with a Structure Sensor Pro and Monocle SS application. The 3D surfaces were manually registered to the CT images and nearest distances between CT surface and 3D surface were calculated. Nearest distance calculations were also performed with a 2 mm CT rind structure that excluded select regions not relevant to surface detection. To demonstrate the advantage of having additional scan length for clearance checks, a collision simulated plan was evaluated with a shortened CT scan length versus with additional 3D surface information.
Results
CT rind to 3D scan surface mean distances for each patient were 1.65 mm, 0.99 mm, −0.93 mm, −0.79 mm, and −1.53 mm with standard deviations of 6.51 mm, 6.71 mm, 3.38 mm, 2.84 mm, and 2.69 mm.
Conclusion
This study demonstrates the feasibility of an accessible low‐cost 3D surface scanning solution to supplement clipped field‐of‐view CT scans and reduce CT scan lengths by obtaining important patient surface data for improved calculation accuracy, treatment clearance checks, and SGRT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, WashU Medicine, 4921 Parkview Place, St. Louis, Missouri, USA.
- Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.
- Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, Florida, USA.
- UF Health Proton Therapy Institute, Jacksonville, Florida, USA.
Zitieren
APA
Jackson, A. M., Shrestha, D. K., & Johnson, P. B. (2025). Three‐dimensional surface scanning of patients to extend computed tomography field of view and scan length for dose calculation and collision avoidance in radiotherapy treatment planning. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70333
Vancouver
Jackson AM, Shrestha DK, Johnson PB. Three‐dimensional surface scanning of patients to extend computed tomography field of view and scan length for dose calculation and collision avoidance in radiotherapy treatment planning. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70333
BibTeX
@article{Jackson2025,
author = {Jackson, Amanda M and Shrestha, Deepak K and Johnson, Perry B},
title = {Three‐dimensional surface scanning of patients to extend computed tomography field of view and scan length for dose calculation and collision avoidance in radiotherapy treatment planning},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70333}
}
EUVIMED