Virtual 3D surface imaging system for the validation of tumor tracking algorithm used in surface‐guided radiotherapy
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70290 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Xiaolong Wu und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Anhui Medical University, Hefei, China. Hauptthemen laut MEDLINE: Algorithms, Radiotherapy, Image-Guided, Imaging, Three-Dimensional, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Shandong Province, Anhui Province Funds for Distinguished Young Scientists.
Abstract (Originalsprache)
Purpose
To develop a surface 3D reconstruction simulation system based on the Phase Measurement Profilometry (PMP) and the Unity physics engine and validate its feasibility for assisting in the development of the Surface‐Guided Radiotherapy (SGRT) tumor tracking algorithm.
Methods
The components, such as cameras and projectors, are set up in the Unity environment to enable structured‐light‐based surface 3D reconstruction simulation using the PMP. This process includes structured light projection, camera calibration, phase unwrapping, and point cloud reconstruction procedures. The influence of parameter settings on the effectiveness of 3D reconstruction is investigated, including different distances and angles between the camera and the measurement surface, as well as variations in light intensity. The simulation capabilities of the system are validated by comparing surface imaging of the same human torso model in a radiotherapy room environment and within the simulation system. Additionally, the simulation system is further utilized to acquire surface imaging data required for the SGRT tumor tracking algorithm. A comparison is made between this data and the idealized skin surface imaging data obtained directly from CT reconstruction segmentation to verify the system's supportive role in the development of the SGRT tumor tracking algorithm.
Results
The effects of varying light intensity and object positioning in the simulation system are consistent with those reported in previous studies conducted in real‐world environments. The root mean square errors (RMSE) of the surface imaging point clouds from different perspectives between the simulation system and the actual environment are 0.46, 0.47, and 0.52 mm, all at sub‐millimeter levels. The validation in the development of the SGRT tumor tracking algorithm indicates that the simulation system enables the SGRT algorithm development to avoid relying on overly idealized surface imaging data.
Conclusions
The simulation system based on PMP and Unity has been proposed, enabling a broader range of measurement conditions to be set in the virtual environment, thereby saving costs for measurements in real‐world scenarios. This system can also be utilized to assist in the validation of the SGRT tumor tracking algorithms, thereby advancing progress in this field.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering, Anhui Medical University, Hefei, China.
- Anhui Province Key Laboratory of Medical Physics and Technology, Institute of Health and Medical Technology, Hefei Institutes of Physical Science, Hefei Cancer Hospital, Chinese Academy of Sciences, Hefei, P.R. China.
- Information Materials and Intelligent Sensing Laboratory of Anhui Province, Anhui University, Hefei, P.R. China.
ORCID-Profile: Ziwen Wei
Förderung
- National Natural Science Foundation of China
- Natural Science Foundation of Shandong Province
- Anhui Province Funds for Distinguished Young Scientists
- Collaborative Innovation Program of Hefei Science Center, CAS
- China Postdoctoral Science Foundation
- Project of Postdoctoral Innovation of Shandong Province
Zitieren
APA
Wu, X., Wei, Z., Zhai, S., Zhang, Y., Liu, Z., Jiang, T., Zhang, L., & Qian, J. (2025). Virtual 3D surface imaging system for the validation of tumor tracking algorithm used in surface‐guided radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70290
Vancouver
Wu X, Wei Z, Zhai S, Zhang Y, Liu Z, Jiang T, et al. Virtual 3D surface imaging system for the validation of tumor tracking algorithm used in surface‐guided radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70290
BibTeX
@article{Wu2025,
author = {Wu, Xiaolong and Wei, Ziwen and Zhai, Shaozhuang and Zhang, Yang and Liu, Zhihua and Jiang, Tao and Zhang, Lei and Qian, Junchao},
title = {Virtual 3D surface imaging system for the validation of tumor tracking algorithm used in surface‐guided radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70290}
}