Improved organ absorbed dose estimation in abdominal CT using DICOM header‐based tube current modulation profiles: Validation with measurements and Monte Carlo simulations
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70321 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Chatnapa Nuntue und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Quantum Medical Technology, Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Ishikawa, …. Hauptthemen laut MEDLINE: Monte Carlo Method, Phantoms, Imaging, Tomography, X-Ray Computed, Radiography, Abdominal. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Precise determination of organ absorbed doses in computed tomography (CT) is vital for evaluating radiation risk and optimizing radiation dose. This study aimed to improve the accuracy of organ absorbed dose estimates in abdominal CT by integrating tube current modulation (TCM) profiles—extracted from the Digital Imaging and Communications in Medicine (DICOM) headers—into Monte Carlo (MC) simulations. The proposed study was validated by comparing the simulation results with direct measurements.
Methods
The Particle and Heavy Ion Transport code System (PHITS), an MC simulation model, was validated by comparing its simulated weighted CT dose index (CTDI w ) with measurements obtained using a 32‐cm PMMA phantom on a 256‐slice multidetector CT scanner. Organ doses were measured using radiophotoluminescence glass dosimeters placed at the anatomical locations of various organs within an abdominal phantom. Scans were performed under both fixed tube current conditions and TCM with noise indices (NI) of 9 and 11. A voxel‐based phantom model was constructed from CT image data, and TCM profiles were extracted from DICOM headers for use in the simulations.
Results
The simulated CTDI w values were within 10% of the measured data. The relative differences between simulated and measured organ absorbed doses for the liver, gallbladder, stomach, spleen, kidneys, and pancreas ranged from −14.0% to 7.87% across both fixed tube current and TCM protocols. Slightly larger deviations were noted for the kidneys under TCM at NI 9. While the average dose differed depending on the NI setting, the dose distribution trend under TCM was comparable to that of fixed tube current scans.
Conclusion
The agreement between simulated and measured organ doses, including CTDI w , was consistent across all protocols. The observed consistency in dose distribution patterns affirms the reliability and applicability of the proposed correction method for accurate and protocol‐specific organ dose estimation in abdominal CT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Quantum Medical Technology, Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.
- Department of Radiological Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumthani, Thailand.
- Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.
- Division of Nuclear Medicine, Department of Radiology, Faculty of Medicine, Khon Kaen University, Khon Kaen, Thailand.
ORCID-Profile: Chatnapa Nuntue, Kosuke Matsubara, Khajonsak Tantiwetchayanon
Zitieren
APA
Nuntue, C., Matsubara, K., Watanabe, S., Fukushima, K., & Tantiwetchayanon, K. (2025). Improved organ absorbed dose estimation in abdominal CT using DICOM header‐based tube current modulation profiles: Validation with measurements and Monte Carlo simulations. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70321
Vancouver
Nuntue C, Matsubara K, Watanabe S, Fukushima K, Tantiwetchayanon K. Improved organ absorbed dose estimation in abdominal CT using DICOM header‐based tube current modulation profiles: Validation with measurements and Monte Carlo simulations. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70321
BibTeX
@article{Nuntue2025,
author = {Nuntue, Chatnapa and Matsubara, Kosuke and Watanabe, Shu and Fukushima, Kotaro and Tantiwetchayanon, Khajonsak},
title = {Improved organ absorbed dose estimation in abdominal CT using DICOM header‐based tube current modulation profiles: Validation with measurements and Monte Carlo simulations},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70321}
}
EUVIMED