Comparative evaluation of normal tissue objective functions in robotic radiosurgery planning for solitary brain tumors
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70318 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Nadine Sonntag und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Stereotactic and Functional Neurosurgery, University Hospital Cologne, Cologne, Germany. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiosurgery, Brain Neoplasms, Organs at Risk. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The normal tissue objective (NTO) is an inverse planning approach in radiosurgery, also available for the CyberKnife system. By employing a model function, it aims to achieve precise control over the global dose fall‐off in healthy tissue. As a novel technique, NTO can serve as an alternative to the established method, which utilizes layered contours around the target to shape dose gradients and enhance conformity, referred to as Auto‐shells in CyberKnife systems.
Purpose
This study compares the dose distribution achieved with NTO and Auto‐shells to evaluate their respective advantages in CyberKnife treatment planning.
Methods
A total of 45 patients with brain tumors—including 15 vestibular schwannomas, 15 meningiomas, and 15 metastases, all of whom had previously been treated using an Auto‐shells‐generated plan, were analyzed. For each case, an alternative NTO‐based plan was generated and compared with its Auto‐shells counterpart. Key treatment parameters—including nodes, beams, total monitor units (MU), treatment time, new conformity index (nCI), gradient index (GI), and dose exposure volumes to healthy brain tissue (V12Gy and V5Gy)—were evaluated.
Results
Both methods resulted in comparable plans across many indices. Significant differences were particularly in terms of healthy brain tissue dose exposure. With the NTO method, V12Gy and V5Gy were reduced by up to 14%, and in the case of meningiomas and metastases, the GI was reduced by up to 7%. The conformity, described by the nCI, was within 2%. No significant difference was observed in MU.
Conclusion
NTO optimization presents a viable option to the Auto‐shells method for CyberKnife treatment of brain tumors. By reducing healthy brain tissue exposure without increasing monitor units, it enhances dose‐sparing efficiency. However, maintaining optimal conformity remains an important issue, highlighting the trade‐offs between precision and tissue preservation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Stereotactic and Functional Neurosurgery, University Hospital Cologne, Cologne, Germany.
- Department of Radiation Oncology, University Hospital Cologne, Cologne, Germany.
ORCID-Profile: Stefan Hunsche
Zitieren
APA
Sonntag, N., Eichner, M., Judge, M., Kocher, M., Ruess, D., Ruge, M. I., & Hunsche, S. (2025). Comparative evaluation of normal tissue objective functions in robotic radiosurgery planning for solitary brain tumors. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70318
Vancouver
Sonntag N, Eichner M, Judge M, Kocher M, Ruess D, Ruge MI, et al. Comparative evaluation of normal tissue objective functions in robotic radiosurgery planning for solitary brain tumors. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70318
BibTeX
@article{Sonntag2025,
author = {Sonntag, Nadine and Eichner, Markus and Judge, Michael and Kocher, Martin and Ruess, Daniel and Ruge, Maximilian I and Hunsche, Stefan},
title = {Comparative evaluation of normal tissue objective functions in robotic radiosurgery planning for solitary brain tumors},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70318}
}
EUVIMED