Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon

Nicholas Nelson iD, Courtney Oare, Geoff Nelson iD, Thomas Martin, Jessica Huang, Hui Zhao

Journal of Applied Clinical Medical Physics · 2025-09 · 26 (9)

DOI: 10.1002/acm2.70245 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 09.09.2025. Verfasst von Nicholas Nelson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The development of on‐board cone‐beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on‐board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone‐beam computed tomography (CBCT) on the Halcyon platform relative to fan‐beam CT‐based dose calculations and to perform end‐to‐end testing of HyperSight CBCT‐only based treatment planning.

Methods

A HU to mass density curve was measured for the HyperSight CBCT system and implemented into the Eclipse treatment planning system. Following this, computational dosimetric analysis was performed between dose distributions calculated on CT simulation (CT sim ) and HyperSight CBCT images on two anthropomorphic phantoms for pelvic and head and neck treatment sites. Additionally, an end‐to‐end test was carried out for a head and neck intensity modulated radiation therapy (IMRT) plan.

Results

The HU to mass density curves acquired on CT sim and HyperSight CBCT were similar (< 30 HU) for near‐water equivalent materials, but deviated for high‐density materials, with a maximum difference of 150 HU. For dose calculations, excellent agreement between dose calculations performed on CT sim and HyperSight CBCT phantom images was observed, where three‐dimensional gamma pass rates between the two dose distributions were observed to be 90% at 1%/1 mm (5% threshold). For the end‐to‐end test, absolute doses were verified to within 1% of ionization chamber measurements, while Delta4+ and portal dosimetry measurements yielded passing results (gamma pass rate 90%) down to 2%/2 mm criterion.

Conclusion

In this study, the accuracy of dose calculations performed on HyperSight CBCT was found to be within 1% of CT sim calculations for pelvic and head and neck treatment sites. End‐to‐end results using the RANDO anthropomorphic phantom indicate that HyperSight CBCT images are suitable for radiation treatment planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedHead and Neck NeoplasmsImage Processing, Computer-AssistedHumansRadiotherapy DosageBenchmarkingFeasibility StudiesOrgans at Risk

Stichwörter der Autor:innen

HalcyonHyperSightadaptive therapycone‐beam computed tomography (CBCT)dose calculationsdosimetryphoton therapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.

ORCID-Profile: Nicholas Nelson, Geoff Nelson

Zitieren

APA

Nelson, N., Oare, C., Nelson, G., Martin, T., Huang, J., & Zhao, H. (2025). Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon. Journal of Applied Clinical Medical Physics, 26(9). https://doi.org/10.1002/acm2.70245

Vancouver

Nelson N, Oare C, Nelson G, Martin T, Huang J, Zhao H. Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon. Journal of Applied Clinical Medical Physics. 2025;26(9). doi:10.1002/acm2.70245

BibTeX
@article{Nelson2025,
  author = {Nelson, Nicholas and Oare, Courtney and Nelson, Geoff and Martin, Thomas and Huang, Jessica and Zhao, Hui},
  title = {Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {9},
  doi = {10.1002/acm2.70245}
}
Kostenlos starten →