EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Evaluation of Dose Calculation Based on Cone-Beam CT Using Different Measuring Correction Methods for Head and Neck Cancer Patients

Hanshun Gong iD, Bo Liu, Gaolong Zhang, Xiangkun Dai, Baolin Qu, Boning Cai, Chuanbin Xie, Shouping Xu iD

Technology in Cancer Research & Treatment · 2023

DOI: 10.1177/15330338221148317 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Technology in Cancer Research & Treatment, 2023. Verfasst von Hanshun Gong und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, 104607The First Medical Center of PLA General Hospital, Beijing, China. Hauptthemen laut MEDLINE: Head and Neck Neoplasms, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Key R&D Program, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate and compare 2 cone-beam computed tomography (CBCT) correction methods for CBCT-based dose calculation. Materials and Methods: Routine CBCT image sets of 12 head and neck cancer patients who received volumetric modulated arc therapy (VMAT) treatment were retrospectively analyzed. The CBCT images obtained using an on-board imager (OBI) at the first treatment fraction were firstly deformable registered and padded with the kVCT images to provide enough anatomical information about the tissues for dose calculation. Then, 2 CBCT correction methods were developed and applied to correct CBCT Hounsfield unit (HU) values. One method (HD method) is based on protocol-specific CBCT HU to physical density (HD) curve, and the other method (HM method) is based on histogram matching (HM) of HU value. The corrected CBCT images (CBCT HD and CBCT HM for HD and HM methods) were imported into the original planning system for dose calculation based on the HD curve of kVCT (the planning CT). The dose computation result was analyzed and discussed to compare these 2 CBCT-correction methods.

Results

Dosimetric parameters, such as the D mean , D max and D 5% of the target volume in CBCT plan doses, were higher than those in the kVCT plan doses; however, the deviations were less than 2%. The D 2% , in parallel organs such as the parotid glands, the deviations from the CBCT HM plan dose were less than those of the CBCT HD plan dose. The differences were statistically significant ( P < .05). Meanwhile, the V 30 value based on the HM method was better than that based on the HD method in the oral cavity region ( P = .016). In addition, we also compared the γ passing rates of kVCT plan doses with the 2 CBCT plan doses, and negligible differences were found.

Conclusion

The HM method was more suitable for head and neck cancer patients than the HD one. Furthermore, with the CBCT HM -based method, the dose calculation result better matches the kVCT-based dose calculation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Head and Neck NeoplasmsRadiotherapy, Intensity-ModulatedHumansRadiotherapy DosageRetrospective StudiesRadiotherapy Planning, Computer-AssistedPhantoms, ImagingCone-Beam Computed TomographyImage Processing, Computer-Assisted

Stichwörter der Autor:innen

HU-density mapHounsfield unit (HU)adaptive radiation therapycone-beam computed tomographydeformable image registrationhistogram matching algorithm

Autor:innen und Institutionen

  • Department of Radiation Oncology, 104607The First Medical Center of PLA General Hospital, Beijing, China.
  • School of Astronautics, 535276Beihang University, Beijing, China.
  • School of Physics, 535276Beihang University, Beijing, China.
  • Department of Radiation Oncology, National Cancer Center/Cancer Hospital, 12501Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
  • National Cancer Center/National Clinical Research Center for Cancer/Hebei Cancer Hospital, 12501Chinese Academy of Medical Sciences, Langfang, China.

ORCID-Profile: Hanshun Gong, Shouping Xu

Förderung

  • National Key R&D Program
  • National Natural Science Foundation of China

Zitieren

APA

Gong, H., Liu, B., Zhang, G., Dai, X., Qu, B., Cai, B., Xie, C., & Xu, S. (2023). Evaluation of Dose Calculation Based on Cone-Beam CT Using Different Measuring Correction Methods for Head and Neck Cancer Patients. Technology in Cancer Research & Treatment. https://doi.org/10.1177/15330338221148317

Vancouver

Gong H, Liu B, Zhang G, Dai X, Qu B, Cai B, et al. Evaluation of Dose Calculation Based on Cone-Beam CT Using Different Measuring Correction Methods for Head and Neck Cancer Patients. Technology in Cancer Research & Treatment. 2023. doi:10.1177/15330338221148317

BibTeX
@article{Gong2023,
  author = {Gong, Hanshun and Liu, Bo and Zhang, Gaolong and Dai, Xiangkun and Qu, Baolin and Cai, Boning and Xie, Chuanbin and Xu, Shouping},
  title = {Evaluation of Dose Calculation Based on Cone-Beam CT Using Different Measuring Correction Methods for Head and Neck Cancer Patients},
  journal = {Technology in Cancer Research & Treatment},
  year = {2023},
  doi = {10.1177/15330338221148317}
}
Kostenlos starten →