Multi-Institutional Validation of a Practical, Reconfigurable Phantom for End-to-End Verification of Cone Beam CT-Based Online Adaptive Radiation Therapy Systems

Adam D Yock, Annie Cooney, Ivy Cornelison, Sarah Masick, Dennis N Stanley, Xenia Ray

Practical Radiation Oncology · 2026

DOI: 10.1016/j.prro.2026.03.007 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Practical Radiation Oncology, 2026. Verfasst von Adam D Yock und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Vanderbilt University Medical Center, Nashville, Tennessee. Electronic address: adam.d.yock@vumc.org. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Vanderbilt Cancer Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Clinical implementation of online adaptive radiation therapy (ART) requires new quality assurance methods and equipment. We introduce a novel quality assurance phantom designed to verify online ART treatment planning and delivery through practical, comprehensive end-to-end tests. Furthermore, we present the results of a multi-institutional validation of the phantom applied to cone beam computed tomography-guided online ART systems. METHODS AND MATERIALS: The phantom comprises swappable plastic pieces and features a rigid-yet-reconfigurable design to achieve multiple target morphologies with known geometric relationships. The phantom was circulated among 3 institutions. Each institution created 2 treatment plans based on shared dosimetric criteria. For each plan, online ART treatment sessions were conducted, and the plans were delivered to the phantom in multiple geometric configurations. Several treatment plan metrics (planning target volume V100% and D98%; organ-at-risk [OAR] D0.03cc and D2cc) were compared between the original and adaptive plans. Physical dose measurements were made in several locations using either plastic scintillators or an ion chamber, and 4 sessions were analyzed using film. Lastly, target volumes were contoured repeatedly by 3 separate users to estimate intrauser/interuser contouring variability. RESULTS: All 3 institutions used the phantom to test their online ART system's ability to adapt to changing target morphology. Across both plans, changes in target coverage metrics were small (95% of values ≤3.3%) and similar between original and adapted plans. However, changes in morphology increased OAR metrics by 20 to 60 cGy when using the original plan. Adapted plans did not increase dose to the OAR or compromise target coverage. The accurate delivery of the plans was confirmed with >97% of physical measurements ≤5% of their expected value and film gamma rates of 90.7% to 95.9%. Intrauser/interuser target volume variability was 1.0% to 1.7% and 1.3% to 2.4%, respectively. CONCLUSIONS: The phantom was successfully used at 3 separate institutions to verify the entire online ART treatment planning and delivery workflow.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy, Image-GuidedHumansRadiotherapy DosageQuality Assurance, Health CareRadiotherapy, Intensity-Modulated

Autor:innen und Institutionen

  • Department of Radiation Oncology, Vanderbilt University Medical Center, Nashville, Tennessee. Electronic address: adam.d.yock@vumc.org.
  • Department of Radiation Oncology, Vanderbilt University Medical Center, Nashville, Tennessee.
  • Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama.
  • Department of Radiation Medicine and Applied Sciences, University of California San Diego, La Jolla, California.

Förderung

  • Vanderbilt Cancer Center

Zitieren

APA

Yock, A. D., Cooney, A., Cornelison, I., Masick, S., Stanley, D. N., & Ray, X. (2026). Multi-Institutional Validation of a Practical, Reconfigurable Phantom for End-to-End Verification of Cone Beam CT-Based Online Adaptive Radiation Therapy Systems. Practical Radiation Oncology. https://doi.org/10.1016/j.prro.2026.03.007

Vancouver

Yock AD, Cooney A, Cornelison I, Masick S, Stanley DN, Ray X. Multi-Institutional Validation of a Practical, Reconfigurable Phantom for End-to-End Verification of Cone Beam CT-Based Online Adaptive Radiation Therapy Systems. Practical Radiation Oncology. 2026. doi:10.1016/j.prro.2026.03.007

BibTeX
@article{Yock2026,
  author = {Yock, Adam D and Cooney, Annie and Cornelison, Ivy and Masick, Sarah and Stanley, Dennis N and Ray, Xenia},
  title = {Multi-Institutional Validation of a Practical, Reconfigurable Phantom for End-to-End Verification of Cone Beam CT-Based Online Adaptive Radiation Therapy Systems},
  journal = {Practical Radiation Oncology},
  year = {2026},
  doi = {10.1016/j.prro.2026.03.007}
}
Kostenlos starten →