Univariate and multivariate analysis of the styrofoam fixation device on patient setup errors in radiotherapy

Jun Chen, Yi Guo iD, Changsheng Wang, Simin Lin, Linglong Shao, Linzhen Lan, Feibao Guo

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70181 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Jun Chen und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, Cancer Center, The First Affiliated Hospital, Fujian Medical University, Fuzhou, China. Hauptthemen laut MEDLINE: Radiotherapy Setup Errors, Radiotherapy Planning, Computer-Assisted, Immobilization, Head and Neck Neoplasms. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fujian Provincial Health Technology Project, the Leading Project Foundation of Science and Technology, Fujian Province of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Styrofoam is a patient‐specific immobilization device in radiotherapy; most previous studies about the impact of styrofoam on setup errors have only analyzed a single tumor type, and have not considered the influence of patient's physical condition on the setup errors of styrofoam.

Purpose

This study aims to evaluate the impact of styrofoam device on setup errors in radiotherapy and explore which patient population is more suitable for styrofoam immobilization.

Methods

Univariate and multivariate analyses were conducted to compare the setup errors between the experimental group (styrofoam combined with thermoplastic mask) and the control group (thermoplastic mask alone). All cases were categorized based on tumor location into head and neck, thorax, abdomen, and limb cases, with age, gender, surgical history, educational level, and body mass index (BMI) serving as variables in the multivariate analysis. Intragroup analysis was also performed.

Results

For all included cases, the experimental group had median setup errors of 1.20 , 2.00 , and 1.30 mm in the vertical, longitudinal, and lateral directions, respectively. In contrast, the control group had median setup errors of 1.50 , 2.00 , and 1.85 mm in the same respective directions. Notably, the experimental group demonstrated statistically significant reductions in average setup errors in the longitudinal direction (2.00 vs. 2.87 mm, p < 0.01) and lateral direction (1.90 vs. 2.24 mm, p < 0.01) compared to the control group. The intragroup analysis results indicated that factors such as age, gender, surgical history, and educational level had no significant impact on the setup errors in the experimental group.

Conclusion

The styrofoam fixation device for patient immobilization can effectively reduce setup errors in both the longitudinal and lateral directions, and the styrofoam fixation device is suitable for most people.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Setup ErrorsRadiotherapy Planning, Computer-AssistedImmobilizationHead and Neck NeoplasmsNeoplasmsPatient PositioningHumansMaleFemaleMiddle AgedRadiotherapy DosageMultivariate AnalysisRadiotherapy, Intensity-ModulatedAgedAdult

Stichwörter der Autor:innen

Styrofoamradiotherapysetup errors

Autor:innen und Institutionen

  • Department of Radiotherapy, Cancer Center, The First Affiliated Hospital, Fujian Medical University, Fuzhou, China.
  • Department of Radiotherapy, National Regional Medical Center, Binhai Campus of the First Affiliated Hospital, Fujian Medical University, Fuzhou, China.
  • Key Laboratory of Radiation Biology of Fujian Higher Education Institutions, the First Affiliated Hospital, Fujian Medical University, Fuzhou, China.
  • Department of Radiology, The First Affiliated Hospital, Fujian Medical University, Fuzhou, China.
  • Department of Radiology, National Regional Medical Center, Binhai Campus of the First Affiliated Hospital, Fujian Medical University, Fuzhou, China.

ORCID-Profile: Yi Guo

Förderung

  • Fujian Provincial Health Technology Project
  • the Leading Project Foundation of Science and Technology, Fujian Province of China

Zitieren

APA

Chen, J., Guo, Y., Wang, C., Lin, S., Shao, L., Lan, L., & Guo, F. (2025). Univariate and multivariate analysis of the styrofoam fixation device on patient setup errors in radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70181

Vancouver

Chen J, Guo Y, Wang C, Lin S, Shao L, Lan L, et al. Univariate and multivariate analysis of the styrofoam fixation device on patient setup errors in radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70181

BibTeX
@article{Chen2025,
  author = {Chen, Jun and Guo, Yi and Wang, Changsheng and Lin, Simin and Shao, Linglong and Lan, Linzhen and Guo, Feibao},
  title = {Univariate and multivariate analysis of the styrofoam fixation device on patient setup errors in radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70181}
}
Kostenlos starten →