EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Analysis of Surface Guidance Versus Laser Alignment for Precision Lung Cancer Particle Therapy

Xiyu Zhang iD, Yuze Yang, Jingfang Mao, Yinxiangzi Sheng iD

Technology in Cancer Research & Treatment · 2026

DOI: 10.1177/15330338261425319 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Technology in Cancer Research & Treatment, 2026. Verfasst von Xiyu Zhang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, China. Hauptthemen laut MEDLINE: Lung Neoplasms, Radiotherapy, Image-Guided, Radiotherapy Planning, Computer-Assisted, Patient Positioning. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Shanghai Municipal Health Commission, Shanghai Pujiang Program, Noncommunicable Chronic Diseases-National Science and Technology Major Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Lung particle therapy using pencil beam scanning achieves high dose conformity but remains vulnerable to geometric uncertainties from suboptimal initial setup. Surface-guided radiotherapy (SGRT) improves setup reproducibility in photon workflows, yet evidence in lung particle therapy remains limited. This study evaluates the clinical value of SGRT in improving six degrees of freedom (6-DOF) setup reproducibility in lung cancer particle therapy.

Methods

This retrospective cohort study analyzed 63 lung cancer patients receiving 1277 treatment fractions at our center from February 2023 to January 2024. Comparisons were made between conventional laser-based positioning, which included 983 fractions, and SGRT workflows, which included 294 fractions. Following patient positioning, therapists manually registered orthogonal kilovoltage (kV) x-ray images with planning digitally reconstructed radiographs (DRRs) to calculate 6-DOF correction parameters, including translational (lateral, longitudinal, vertical) and rotational (pitch, roll, yaw) components, and to quantify the pre-correction setup error . Absolute 6-DOF displacements and three-dimensional vector magnitudes (MAG) were measured. The analysis included 36 supine patients with 739 treatment fractions and 27 prone patients with 538 fractions.

Results

The SGRT group exhibited statistically significant reductions in median shifts for lateral (0.25 cm to 0.21 cm, p = 0.021), longitudinal (0.25 cm to 0.21 cm, p = 0.014), pitch (1.0° to 0.8°, p = 0.001), and MAG (0.59 cm to 0.53 cm, p = 0.002) compared to conventional methods. These improvements in median values were more pronounced in supine-positioned patients, while no significant differences were observed in prone-positioned patients. Furthermore, substantial reductions were achieved in ninth decile deviations (1.09 cm to 1.03 cm), and the third quartile deviations (0.83 cm to 0.74 cm) in the overall cohort.

Conclusion

SGRT enhances setup precision for proton and carbon ion lung cancer radiotherapy, reduces pre-correction setup error, and provides clinical support for patient setup reproducibility.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lung NeoplasmsRadiotherapy, Image-GuidedRadiotherapy Planning, Computer-AssistedPatient PositioningLasersHumansRetrospective StudiesFemaleMaleAgedMiddle AgedRadiotherapy DosageRadiotherapy Setup Errors

Stichwörter der Autor:innen

lung canceroptical surface imaging (OSI)particle therapysetup errorsurface-guided radiotherapy (SGRT)

Autor:innen und Institutionen

  • Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, China.
  • Shanghai Key Laboratory of Radiation Oncology, Shanghai, China.
  • Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai, China.
  • Department of Radiation Oncology, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai 201315, China.

ORCID-Profile: Xiyu Zhang, Yinxiangzi Sheng

Förderung

  • Shanghai Municipal Health Commission
  • Shanghai Pujiang Program
  • Noncommunicable Chronic Diseases-National Science and Technology Major Project

Zitieren

APA

Zhang, X., Yang, Y., Mao, J., & Sheng, Y. (2026). Analysis of Surface Guidance Versus Laser Alignment for Precision Lung Cancer Particle Therapy. Technology in Cancer Research & Treatment. https://doi.org/10.1177/15330338261425319

Vancouver

Zhang X, Yang Y, Mao J, Sheng Y. Analysis of Surface Guidance Versus Laser Alignment for Precision Lung Cancer Particle Therapy. Technology in Cancer Research & Treatment. 2026. doi:10.1177/15330338261425319

BibTeX
@article{Zhang2026,
  author = {Zhang, Xiyu and Yang, Yuze and Mao, Jingfang and Sheng, Yinxiangzi},
  title = {Analysis of Surface Guidance Versus Laser Alignment for Precision Lung Cancer Particle Therapy},
  journal = {Technology in Cancer Research & Treatment},
  year = {2026},
  doi = {10.1177/15330338261425319}
}
Kostenlos starten →