EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone‐beam CT in rectal cancer radiotherapy

Junjie Ruan, Xiaotong Huang, Tong Wang, Xiuying Mai, Chuyan Lin, Fanghua Li, Yunfeng Li, Feng Chi, Bin Li iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.14573 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Junjie Ruan und 8 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Oncology in South China, Guangdong Provincial Clinical Research Center for Cancer, Sun Yat-sen University Cancer Center, Guan…. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Rectal Neoplasms, Radiotherapy Planning, Computer-Assisted, Immobilization. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Guangdong Basic and Applied Basic Research Foundation, Special Scientific Research Fund for Medical Physics & Technology of Guangdong Biomedical Engineering Society.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The objective of this study is to evaluate the impact of different belly board and daily changes in patient's body‐mass factor (BMF) on setup displacement in radiotherapy for rectal cancer.

Methods

Twenty‐five patients were immobilized using the thermoplastic mask with belly board (TM‐BB), and 30 used the vacuum bag cushion with belly board (VBC‐BB), performing daily cone‐beam computed tomography (CBCT) scans 625 times and 750 times, respectively. Daily pretreatment CBCT scans were registered to the planned CT images for BMF change determination and setup displacement measurement. Independent t‐tests compared setup displacement between the two groups in left‐right (LR), superior‐inferior (SI), and anterior‐posterior (AP) directions, as well as the BMF changes. The impact of daily BMF changes on setup displacement was evaluated using multivariate logistic regression and 10‐fold cross‐validation.

Results

The setup displacement for TM‐BB in the LR, SI, and AP directions were 0.31 ± 0.25, 0.58 ± 0.40, and 0.19 ± 0.18 cm, respectively, while VBC‐BB showed 0.19 ± 0.15, 0.26 ± 0.22, and 0.36 ± 0.29 cm in the corresponding directions, respectively. Margins of planning target volume (PTV) for TM‐BB were 8, 10, and 6 mm in LR, SI, and AP directions, while VBC‐BB showed margins of 5,7, and 8 mm, respectively. The daily BMF changes for both groups were ranked in descending order as follows: sacral rotation angle (RS), hip lateral diameter (HLD), and hip anterior‐posterior diameter (HAPD). HAPD was the main factor affecting setup displacement in both the AP and SI directions in TM‐BB, while RS was the primary factor for setup displacement in the AP direction in VBC‐BB.

Conclusion

Compared with TM‐BB, VBC‐BB had a larger AP displacement but smaller in LR and SI displacement. Daily changes in BMF have distinct effects on setup displacement in different immobilization devices. Image‐guided radiation therapy (IGRT) is highly recommended and BMF changes should be given consideration during radiotherapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyRectal NeoplasmsRadiotherapy Planning, Computer-AssistedImmobilizationRadiotherapy Setup ErrorsPatient PositioningHumansMaleRadiotherapy DosageFemaleRadiotherapy, Intensity-ModulatedMiddle AgedAgedBody Mass IndexImage Processing, Computer-AssistedOrgans at Risk

Stichwörter der Autor:innen

belly board immobilizationbody mass factorcone‐beam CTrectal cancer radiotherapysetup displacement

Autor:innen und Institutionen

  • State Key Laboratory of Oncology in South China, Guangdong Provincial Clinical Research Center for Cancer, Sun Yat-sen University Cancer Center, Guangzhou, P. R. China.

ORCID-Profile: Bin Li

Förderung

  • Guangdong Basic and Applied Basic Research Foundation
  • Special Scientific Research Fund for Medical Physics & Technology of Guangdong Biomedical Engineering Society

Zitieren

APA

Ruan, J., Huang, X., Wang, T., Mai, X., Lin, C., Li, F., Li, Y., Chi, F., & Li, B. (2025). Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone‐beam CT in rectal cancer radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14573

Vancouver

Ruan J, Huang X, Wang T, Mai X, Lin C, Li F, et al. Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone‐beam CT in rectal cancer radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14573

BibTeX
@article{Ruan2025,
  author = {Ruan, Junjie and Huang, Xiaotong and Wang, Tong and Mai, Xiuying and Lin, Chuyan and Li, Fanghua and Li, Yunfeng and Chi, Feng and Li, Bin},
  title = {Impact of belly board immobilization devices and body mass factor on setup displacement using daily cone‐beam CT in rectal cancer radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.14573}
}
Kostenlos starten →