Long‐term stability analysis of beam shape in a robotic radiosurgery system
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70123 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Ryoichi Hinoto und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Saitama Red Cross Hospital, Saitama, Japan. Hauptthemen laut MEDLINE: Radiosurgery, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Robotic Surgical Procedures. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study aimed to investigate the long‐term stability of CyberKnife beam profile parameters and assess their compliance with existing quality assurance (QA) guidelines. We evaluated beam profiles in both standard and diagonal planes over 3.5 years post‐installation to detect potential issues and ensure consistent beam quality. The findings will contribute to validating and refining current QA practices for CyberKnife systems.
Methods
Beam profile measurements were performed monthly using an Octavius 1000SRS detector array. The profiles were evaluated in terms of the beam shape constancy within 2%, and the penumbra, symmetry, and flatness were analyzed using statistical process control methods. Temporal changes in the dose profiles were visualized using dose difference heat maps. The relationship between the beam parameters and accumulated monitor units was also investigated.
Results
The 2% profile constancy check accurately detected magnetron deterioration 2 months before failure, confirming its high sensitivity for beam stability monitoring. While symmetry and flatness remained within 0.7% throughout over 100 × 10 6 monitor units of operation, penumbra exhibited greater responsiveness to magnetron‐induced changes but did not consistently flag all orientations. Additionally, statistical analyses and heat maps revealed gradual profile shifts independent of acute component failures, highlighting the importance of multifaceted QA strategies.
Conclusions
These findings reinforce the effectiveness of the 2% profile constancy check for early detection of magnetron failure and support its adoption in current CyberKnife guidelines. At the same time, symmetry, flatness, and penumbra parameters remain valuable for characterizing gradual profile variations. Collectively, this study underscores the need for comprehensive beam monitoring and regular maintenance to sustain optimal CyberKnife system performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Saitama Red Cross Hospital, Saitama, Japan.
- Department of Radiation Oncology, Saitama Municipal Hospital, Saitama, Japan.
- Faculty of Medicine, University of Tsukuba, Tsukuba, Japan.
ORCID-Profile: Ryoichi Hinoto, Shiho Kashiyama
Zitieren
APA
Hinoto, R., Kashiyama, S., Eriguchi, T., Tsukamoto, N., & Sakae, T. (2025). Long‐term stability analysis of beam shape in a robotic radiosurgery system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70123
Vancouver
Hinoto R, Kashiyama S, Eriguchi T, Tsukamoto N, Sakae T. Long‐term stability analysis of beam shape in a robotic radiosurgery system. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70123
BibTeX
@article{Hinoto2025,
author = {Hinoto, Ryoichi and Kashiyama, Shiho and Eriguchi, Takahisa and Tsukamoto, Nobuhiro and Sakae, Takeji},
title = {Long‐term stability analysis of beam shape in a robotic radiosurgery system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70123}
}
EUVIMED