Long‐term stability analysis of beam shape in a robotic radiosurgery system

Ryoichi Hinoto iD, Shiho Kashiyama iD, Takahisa Eriguchi, Nobuhiro Tsukamoto, Takeji Sakae

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70123 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Ryoichi Hinoto und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Saitama Red Cross Hospital, Saitama, Japan. Hauptthemen laut MEDLINE: Radiosurgery, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Robotic Surgical Procedures. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to investigate the long‐term stability of CyberKnife beam profile parameters and assess their compliance with existing quality assurance (QA) guidelines. We evaluated beam profiles in both standard and diagonal planes over 3.5 years post‐installation to detect potential issues and ensure consistent beam quality. The findings will contribute to validating and refining current QA practices for CyberKnife systems.

Methods

Beam profile measurements were performed monthly using an Octavius 1000SRS detector array. The profiles were evaluated in terms of the beam shape constancy within 2%, and the penumbra, symmetry, and flatness were analyzed using statistical process control methods. Temporal changes in the dose profiles were visualized using dose difference heat maps. The relationship between the beam parameters and accumulated monitor units was also investigated.

Results

The 2% profile constancy check accurately detected magnetron deterioration 2 months before failure, confirming its high sensitivity for beam stability monitoring. While symmetry and flatness remained within 0.7% throughout over 100 × 10 6 monitor units of operation, penumbra exhibited greater responsiveness to magnetron‐induced changes but did not consistently flag all orientations. Additionally, statistical analyses and heat maps revealed gradual profile shifts independent of acute component failures, highlighting the importance of multifaceted QA strategies.

Conclusions

These findings reinforce the effectiveness of the 2% profile constancy check for early detection of magnetron failure and support its adoption in current CyberKnife guidelines. At the same time, symmetry, flatness, and penumbra parameters remain valuable for characterizing gradual profile variations. Collectively, this study underscores the need for comprehensive beam monitoring and regular maintenance to sustain optimal CyberKnife system performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedRobotic Surgical ProceduresPhantoms, ImagingHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

CyberKnifeflatnesslong‐term analysispenumbrasymmetry

Autor:innen und Institutionen

  • Department of Radiation Oncology, Saitama Red Cross Hospital, Saitama, Japan.
  • Department of Radiation Oncology, Saitama Municipal Hospital, Saitama, Japan.
  • Faculty of Medicine, University of Tsukuba, Tsukuba, Japan.

ORCID-Profile: Ryoichi Hinoto, Shiho Kashiyama

Zitieren

APA

Hinoto, R., Kashiyama, S., Eriguchi, T., Tsukamoto, N., & Sakae, T. (2025). Long‐term stability analysis of beam shape in a robotic radiosurgery system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70123

Vancouver

Hinoto R, Kashiyama S, Eriguchi T, Tsukamoto N, Sakae T. Long‐term stability analysis of beam shape in a robotic radiosurgery system. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70123

BibTeX
@article{Hinoto2025,
  author = {Hinoto, Ryoichi and Kashiyama, Shiho and Eriguchi, Takahisa and Tsukamoto, Nobuhiro and Sakae, Takeji},
  title = {Long‐term stability analysis of beam shape in a robotic radiosurgery system},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70123}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche