Monte Carlo modeling of radiation dose from radiation therapy with superficial x‐rays
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70062 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Reham Barghash und 3 weiteren Autor:innen. Beteiligte Einrichtung: Environmental and Radiological Health Sciences, Colorado State University, Fort Collins, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Organs at Risk. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Superficial x‐rays (50–100 kVp) are used for treating non‐melanoma skin cancer and intraoperative radiation therapy (IORT). At these energies, the photoelectric effect significantly increases absorbed dose to bone compared to soft tissue.
Methods
We used EGSnrc MC simulations to investigate bone dose enhancement during radiotherapy with the Sensus SRT‐100 machine. Simulated beams were validated against laboratory measurements and compared to a commercial treatment planning system (SmART‐ATP). Transmission simulations indirectly predicted bone dosage. Simulated beams were utilized as a mock treatment plan from a human cone‐beam computed tomography (CBCT) image.
Results
EGSnrc accurately modeled the Sensus SRT‐100 beams (100, 70, and 50 kVp) with a root mean square error (RMSE) of percentage depth dose ratios between Monte Carlo predictions and lab measurements of 1.66, 0.47, and 0.99, respectively. PDDs from simulations of a water phantom with bone slabs showed peak doses at water–bone interfaces relative to surface doses. At 0.3 cm depth bone slab, doses reached 410%, 490%, and 510% for 50, 70, and 100 kVp, respectively. At 1.5 cm depth, doses were 140%, 215%, and 270%. At 2.5 cm depth, peak doses were 74%, 130%, and 170% for 50, 70, and 100 kVp, respectively. A simulated treatment plan (4 Gy surface dose) using a CBCT of a human head predicted the dose to the skull to be around 20, 19, and 15 Gy for the 100, 70, and 50 kVp beams, respectively.
Conclusions
The study demonstrated EGSnrc's efficiency in conjunction with SmART‐ATP for treatment planning. MC simulations effectively quantified bone dose enhancement during superficial x‐ray radiotherapy, highlighting its importance in treatment planning and dose calculations. Clinicians should consider measuring bone depth prior to treatment to avoid excessive bone dose.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Environmental and Radiological Health Sciences, Colorado State University, Fort Collins, USA.
- College of Applied Medical Sciences, King Saud bin Abdulaziz University for Health Sciences, Riyadh, Saudi Arabia.
- King Abdullah International Medical Research Center, King Saud bin Abdulaziz University for Health Sciences, Riyadh, Saudi Arabia.
- Department of Clinical Sciences, Colorado State University, Fort Collins, USA.
- Veterinary Teaching Hospital, Colorado State University, Fort Collins, USA.
ORCID-Profile: Amber R Prebble
Zitieren
APA
Barghash, R., Martin, T. W., Prebble, A. R., & Leary, D. (2025). Monte Carlo modeling of radiation dose from radiation therapy with superficial x‐rays. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70062
Vancouver
Barghash R, Martin TW, Prebble AR, Leary D. Monte Carlo modeling of radiation dose from radiation therapy with superficial x‐rays. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70062
BibTeX
@article{Barghash2025,
author = {Barghash, Reham and Martin, Tiffany W and Prebble, Amber R and Leary, Del},
title = {Monte Carlo modeling of radiation dose from radiation therapy with superficial x‐rays},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70062}
}