Technical note: A comparison of in‐house 3D‐printed and commercially available patient‐specific skin collimators for use with electron beam therapy
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14366 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Steven M Herchko und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA. Hauptthemen laut MEDLINE: Electrons, Radiotherapy Dosage, Printing, Three-Dimensional, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Skin collimation is a useful tool in electron beam therapy (EBT) to decrease the penumbra at the field edge and minimize dose to nearby superficial organs at risk (OARs), but manually fabricating these collimation devices in the clinic to conform to the patient's anatomy can be a difficult and time intensive process. This work compares two types of patient‐specific skin collimation (in‐house 3D printed and vendor‐provided machined brass) using clinically relevant metrics.
Methods
Attenuation measurements were performed to determine the thickness of each material needed to adequately shield both 6 and 9 MeV electron beams. Relative and absolute dose planes at various depths were measured using radiochromic film to compare the surface dose, flatness, and penumbra of the different skin collimation materials.
Results
Clinically acceptable thicknesses of each material were determined for both 6 and 9 MeV electron beams. Field width, flatness, and penumbra results between the two systems were very similar and significantly improved compared to measurements performed with no surface collimation.
Conclusion
Both skin collimation methods investigated in this work generate sharp penumbras at the field edge and can minimize dose to superficial OARs compared to treatment fields with no surface collimation. The benefits of skin collimation are greatest for lower energy electron beams, and the benefits decrease as the measurement depth increases. Using bolus with skin collimation is recommended to avoid surface dose enhancement seen with collimators placed on the skin surface. Ultimately, the appropriate choice of material will depend on the desire to create these devices in‐house or outsource the fabrication to a vendor.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.
ORCID-Profile: Steven M Herchko, Sridhar Yaddanapudi
Zitieren
APA
Herchko, S. M., Rutenberg, M. S., Beltran, C. J., & Yaddanapudi, S. (2024). Technical note: A comparison of in‐house 3D‐printed and commercially available patient‐specific skin collimators for use with electron beam therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14366
Vancouver
Herchko SM, Rutenberg MS, Beltran CJ, Yaddanapudi S. Technical note: A comparison of in‐house 3D‐printed and commercially available patient‐specific skin collimators for use with electron beam therapy. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14366
BibTeX
@article{Herchko2024,
author = {Herchko, Steven M and Rutenberg, Michael S and Beltran, Chris J and Yaddanapudi, Sridhar},
title = {Technical note: A comparison of in‐house 3D‐printed and commercially available patient‐specific skin collimators for use with electron beam therapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14366}
}
EUVIMED