EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

High spatial resolution dosimetry for radiation oncology with “MagicPlates,” a new 976‐pixel monolithic silicon detector

Ilia Filipev iD, Saree Alnaghy, Martin Carolan, Jason Paino, Marco Petasecca, Dean Cutajar, Joel Poder iD, Justin B Davies, Bradley M Oborn, Nicholas Hardcastle, Susanna Guatelli, Michael Lerch iD und 2 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70015 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Ilia Filipev und 13 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Radiation Physics, University of Wollongong, New South Wales, Australia. Hauptthemen laut MEDLINE: Silicon, Radiometry, Particle Accelerators, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

We introduce the next generation of “MagicPlate” 2D monolithic pixelated semiconductor detectors – MagicPlate‐976 (MP976). It features a larger array area, higher spatial resolution, and does not require external triggering. We perform a comprehensive characterization for small‐field steep‐dose‐gradient dosimetry applications in radiation therapy focusing on x‐ray beams used in stereotactic treatments.

Methods

The MP976, developed by the Centre for Medical Radiation Physics, consists of 976 ion‐implanted diodes on a thin n‐type epitaxial silicon substrate with a total array area of 58 × 58 mm 2 . The central region has “small” diodes with an area of 0.2 × 0.2 mm 2 and 1 mm pitch and the peripheral region has “large” diodes with an area of 0.6 × 0.6 mm 2 and 2 mm pitch. The detector was primed with 10 kGy (Co‐60) and tested using a Varian TrueBeam linear accelerator for sensitivity change and dose linearity, and variations in response due to dose‐per‐pulse and beam incidence angle. Output factors, depth dose, and beam profiles were measured and compared with reference data.

Results

After the 10 kGy, the sensitivity declined by (74 ± 5)% for “large” diodes and by (78 ± 7)% for the “small” ones, the dose‐per‐pulse (DPP) dependence was in the range of commercially available diodes, however, a difference in the DPP dependence between the “large” and “small” diodes of (8.4 ± 0.2)% was found in the studied DPP range from 0.131–1.111 mGy/pulse. The minimum angular response was at 90° for 6 MV and 100° for 10 MV flattened beams (76% and 82%, respectively). The output factors and depth dose response showed agreement with the reference within 3.1% and 1%, respectively. Deviation in small field 80%/20% penumbra measurements was within 0.5 mm for 6 MV FF and 0.3 mm for 10 MV FFF. Full width at half maximum (FWHM) for the beam profiles agreed within 0.5 mm for both beam qualities.

Conclusion

The new MagicPlate‐976 detector system is shown to be suitable for dosimetry in small fields and steep dose gradients. It provides 1 mm spatial resolution in the central region and 2 mm on the periphery and has no dependence on the field size. The system's high spatial and temporal resolution opens new opportunities for trigger‐less, film‐less, and time‐resolved verification and error identification for complex stereotactic treatment plans.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SiliconRadiometryParticle AcceleratorsRadiotherapy Planning, Computer-AssistedRadiation OncologySemiconductorsNeoplasmsPhantoms, ImagingRadiosurgeryHumansRadiotherapy DosageEquipment Design

Stichwörter der Autor:innen

high spatial resolutionmonolithic 2D arraysilicon pixelated detectorsmall field dosimetrystereotactic radiotherapy

Substanzen

Silicon

Autor:innen und Institutionen

  • Centre for Medical Radiation Physics, University of Wollongong, New South Wales, Australia.
  • Nelune Comprehensive Cancer Centre, Prince of Wales Hospital, New South Wales, Australia.
  • Illawarra Cancer Care Centre, Illawarra Shoalhaven Local Health District, Wollongong, New South Wales, Australia.
  • St George Hospital Cancer Care Centre, Kogarah, New South Wales, Australia.
  • Australian Nuclear Science and Technology Organisation, Lucas Heights, New South Wales, Australia.
  • Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiooncology-OncoRay, Dresden, Germany.

ORCID-Profile: Ilia Filipev, Joel Poder, Michael Lerch, Anatoly Rosenfeld

Zitieren

APA

Filipev, I., Alnaghy, S., Carolan, M., Paino, J., Petasecca, M., Cutajar, D., Poder, J., Davies, J. B., Oborn, B. M., Hardcastle, N., Guatelli, S., Lerch, M., Kron, T., & Rosenfeld, A. (2025). High spatial resolution dosimetry for radiation oncology with “MagicPlates,” a new 976‐pixel monolithic silicon detector. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70015

Vancouver

Filipev I, Alnaghy S, Carolan M, Paino J, Petasecca M, Cutajar D, et al. High spatial resolution dosimetry for radiation oncology with “MagicPlates,” a new 976‐pixel monolithic silicon detector. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70015

BibTeX
@article{Filipev2025,
  author = {Filipev, Ilia and Alnaghy, Saree and Carolan, Martin and Paino, Jason and Petasecca, Marco and Cutajar, Dean and Poder, Joel and Davies, Justin B and Oborn, Bradley M and Hardcastle, Nicholas and Guatelli, Susanna and Lerch, Michael and Kron, Tomas and Rosenfeld, Anatoly},
  title = {High spatial resolution dosimetry for radiation oncology with “MagicPlates,” a new 976‐pixel monolithic silicon detector},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70015}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche