Technical Note: Investigating of dosimetric leaf gap and leaf transmission factor variations across gantry and collimator angles in volumetric modulation arc therapy

Aram Rostami, Mojtaba Barzegar iD, Muhammad Usman, Satheesh Prasad Paloor, Abbass Yousef Mkanna, Alla Fuad Al-Sabahi, Rabih Wafiq Hammoud

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14523 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Aram Rostami und 6 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Department, National Center for Cancer Care and Research, Doha, Qatar. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Phantoms, Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Qatar National Library.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study investigates the influence of gantry and collimator angles on the dosimetric leaf gap (DLG) and leaf transmission factor (LTF) in a Varian LINAC equipped with rounded‐end multi‐leaf collimators (MLCs). While Varian guidelines recommend DLG measurements at zero degrees for both gantry and collimator, this research aims to address the knowledge gap by assessing DLG and LTF variations at different gantry and collimator angles.

Methods

Measurements were conducted using a Varian TrueBeam LINAC with a Millennium 120‐leaf MLC and Eclipse TPS version 16.1. The beams utilized in this study had energies of 6 MV, 10 MV, 6 FFF, and 10 FFF. LTF and DLG were determined using ionization chambers in solid water phantoms at various gantry angles (0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°). For each gantry angle, measurements were also taken at various collimator angles (0°, 45°, 90°, and 315°). Dosimetric impacts were evaluated through VMAT Picket Fence tests and patient‐specific verification using portal dosimetry for 10 clinical VMAT plans.

Results

LTF values showed no significant variation across gantry and collimator angles. However, DLG values exhibited notable differences depending on the gantry angle and were independent of the collimator angle. The highest DLG value was observed at a gantry angle of 270 degrees, while the lowest was at 90 degrees. The AXB DLG Average (averaging seven measurements of DLGs at different gantry angles) model demonstrated the best agreement between measured and calculated dose distributions, indicating the importance of considering averaged DLG values across multiple gantry angles for accurate dose calculations.

Conclusion

Our study highlights the variability of DLG with gantry angle alterations, contrary to Varian guidelines recommending DLG measurements at zero gantry angle only. We advocate for utilizing an averaged DLG value from measurements across multiple gantry angles, as outlined in our methodology.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedPhantoms, ImagingParticle AcceleratorsHumansRadiometryNeoplasmsOrgans at Risk

Stichwörter der Autor:innen

DLGLTFMLCgantry and collimator

Autor:innen und Institutionen

  • Radiation Oncology Department, National Center for Cancer Care and Research, Doha, Qatar.
  • Society for Brain Mapping and Therapeutic, Los Angles, California, USA.
  • Intelligent Quantitative Bio-Medical Imaging (IQBMI), Tehran, Iran.

ORCID-Profile: Mojtaba Barzegar

Förderung

  • Qatar National Library

Zitieren

APA

Rostami, A., Barzegar, M., Usman, M., Paloor, S. P., Mkanna, A. Y., Al-Sabahi, A. F., & Hammoud, R. W. (2024). Technical Note: Investigating of dosimetric leaf gap and leaf transmission factor variations across gantry and collimator angles in volumetric modulation arc therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14523

Vancouver

Rostami A, Barzegar M, Usman M, Paloor SP, Mkanna AY, Al-Sabahi AF, et al. Technical Note: Investigating of dosimetric leaf gap and leaf transmission factor variations across gantry and collimator angles in volumetric modulation arc therapy. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14523

BibTeX
@article{Rostami2024,
  author = {Rostami, Aram and Barzegar, Mojtaba and Usman, Muhammad and Paloor, Satheesh Prasad and Mkanna, Abbass Yousef and Al-Sabahi, Alla Fuad and Hammoud, Rabih Wafiq},
  title = {Technical Note: Investigating of dosimetric leaf gap and leaf transmission factor variations across gantry and collimator angles in volumetric modulation arc therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14523}
}
Kostenlos starten →