Treatment plan complexity quantification for predicting gamma passing rates in patient‐specific quality assurance for stereotactic volumetric modulated arc therapy

Xudong Xue, Shunyao Luan, Yi Ding, Xiangbin Li, Dan Li, Jingya Wang, Chi Ma, Man Jiang, Wei Wei, Xiao Wang iD

Journal of Applied Clinical Medical Physics · 2024-09 · 25 (9)

DOI: 10.1002/acm2.14432 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 18.06.2024. Verfasst von Xudong Xue und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Hubei Cancer Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Quality Assurance, Health Care. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Health Commission of Hubei Province Scientific Research Project, Natural Science Foundation of Hubei Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the beam complexity of stereotactic Volumetric Modulated Arc Therapy (VMAT) plans quantitively and predict gamma passing rates (GPRs) using machine learning.

Methods

The entire dataset is exclusively made of stereotactic VMAT plans (301 plans with 594 beams) from Varian Edge LINAC. The GPRs were analyzed using Varian's portal dosimetry with 2%/2 mm criteria. A total of 27 metrics were calculated to investigate the correlation between metrics and GPRs. Random forest and gradient boosting models were developed and trained to predict the GPRs based on the extracted complexity features. The threshold values of complexity metric were obtained to predict a given beam to pass or fail from ROC curve analysis.

Results

The three moderately significant values of Spearman's rank correlation to GPRs were 0.508 ( p < 0.001), 0.445 ( p < 0.001), and −0.416 ( p < 0.001) for proposed metric LAAM, the ratio of the average aperture area over jaw area (AAJA) and index of modulation, respectively. The random forest method achieved 98.74% prediction accuracy with mean absolute error of 1.23% using five‐fold cross‐validation, and 98.71% with 1.25% for gradient boosting regressor method, respectively. LAAM, leaf travelling distance (LT), AAJA, LT modulation complexity score (LTMCS) and index of modulation, were the top five most important complexity features. The LAAM metric showed the best performance with AUC value of 0.801, and threshold value of 0.365.

Conclusions

The calculated metrics were effective in quantifying the complexity of stereotactic VMAT plans. We have demonstrated that the GPRs could be accurately predicted using machine learning methods based on extracted complexity metrics. The quantification of complexity and machine learning methods have the potential to improve stereotactic treatment planning and identify the failure of QA results promptly.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedRadiotherapy DosageQuality Assurance, Health CareHumansOrgans at RiskMachine LearningRadiosurgeryGamma RaysAlgorithmsParticle Accelerators

Stichwörter der Autor:innen

complexity metricmachine learningquality assurancestereotactic VMAT

Autor:innen und Institutionen

  • Department of Radiation Oncology, Hubei Cancer Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
  • Department of Optoelectronic Engineering, Huazhong University of Science and Technology, Wuhan, China.
  • Department of Radiation Oncology, Rutgers-Cancer Institute of New Jersey, Rutgers-Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA.
  • Department of Nuclear Engineering and Technology, School of Energy and Power Engineering, Huazhong University of Science and Technology, Wuhan, China.

ORCID-Profile: Xiao Wang

Förderung

  • National Natural Science Foundation of China
  • Health Commission of Hubei Province Scientific Research Project
  • Natural Science Foundation of Hubei Province

Zitieren

APA

Xue, X., Luan, S., Ding, Y., Li, X., Li, D., Wang, J., Ma, C., Jiang, M., Wei, W., & Wang, X. (2024). Treatment plan complexity quantification for predicting gamma passing rates in patient‐specific quality assurance for stereotactic volumetric modulated arc therapy. Journal of Applied Clinical Medical Physics, 25(9). https://doi.org/10.1002/acm2.14432

Vancouver

Xue X, Luan S, Ding Y, Li X, Li D, Wang J, et al. Treatment plan complexity quantification for predicting gamma passing rates in patient‐specific quality assurance for stereotactic volumetric modulated arc therapy. Journal of Applied Clinical Medical Physics. 2024;25(9). doi:10.1002/acm2.14432

BibTeX
@article{Xue2024,
  author = {Xue, Xudong and Luan, Shunyao and Ding, Yi and Li, Xiangbin and Li, Dan and Wang, Jingya and Ma, Chi and Jiang, Man and Wei, Wei and Wang, Xiao},
  title = {Treatment plan complexity quantification for predicting gamma passing rates in patient‐specific quality assurance for stereotactic volumetric modulated arc therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {9},
  doi = {10.1002/acm2.14432}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche