Simultaneous high‐pitch multi‐energy CT pulmonary angiography using a dual‐source photon‐counting‐detector CT: A phantom experiment
Journal of Applied Clinical Medical Physics · 2024-11 · 25 (11)
DOI: 10.1002/acm2.14496 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.08.2024. Verfasst von Jelena M Mihailovic und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Pulmonary Embolism, Photons, Computed Tomography Angiography. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
A dual‐source CT system can be operated in a high‐pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi‐energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy‐integrating detector (EID) CT can perform simultaneous ME and high‐pitch acquisition. This phantom study aimed to evaluate the ability of a photon‐counting‐detector (PCD) CT to perform simultaneous high‐pitch and ME imaging for CTPA.
Methods
A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high‐pitch and ME EID‐CT scans and a single high‐pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images.
Results
When there was motion, EID‐CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD‐CT in high‐pitch ME were more stable when there was a motion, with a lower standard deviation than EID‐CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID‐CT, while PCD‐CT high‐pitch ME mode showed good agreement between measurements on static or moving images.
Conclusion
PCD‐CT with combined ME and high‐pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Zitieren
APA
Mihailovic, J. M., Bruesewitz, M. R., Swicklik, J. R., Yalon, M., Rajiah, P. S., Fletcher, J. G., McCollough, C. H., & Yu, L. (2024). Simultaneous high‐pitch multi‐energy CT pulmonary angiography using a dual‐source photon‐counting‐detector CT: A phantom experiment. Journal of Applied Clinical Medical Physics, 25(11). https://doi.org/10.1002/acm2.14496
Vancouver
Mihailovic JM, Bruesewitz MR, Swicklik JR, Yalon M, Rajiah PS, Fletcher JG, et al. Simultaneous high‐pitch multi‐energy CT pulmonary angiography using a dual‐source photon‐counting‐detector CT: A phantom experiment. Journal of Applied Clinical Medical Physics. 2024;25(11). doi:10.1002/acm2.14496
BibTeX
@article{Mihailovic2024,
author = {Mihailovic, Jelena M and Bruesewitz, Michael R and Swicklik, Joseph R and Yalon, Mariana and Rajiah, Prabhakar S and Fletcher, Joel G and McCollough, Cynthia H and Yu, Lifeng},
title = {Simultaneous high‐pitch multi‐energy CT pulmonary angiography using a dual‐source photon‐counting‐detector CT: A phantom experiment},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {11},
doi = {10.1002/acm2.14496}
}
EUVIMED