Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in‐depth assessment of metal artifacts

Bente Konst, Linus Ohlsson, Lilian Henriksson, Mårten Sandstedt, Anders Persson, Tino Ebbers

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14386 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Bente Konst und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden. Hauptthemen laut MEDLINE: Artifacts, Heart-Assist Devices, Phantoms, Imaging, Tomography, X-Ray Computed. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CMIV, Swedish Research Council, Swedish Heart and Lung Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Photon counting CT (PCCT) holds promise for mitigating metal artifacts and can produce virtual mono‐energetic images (VMI), while maintaining temporal resolution, making it a valuable tool for characterizing the heart. This study aimed to evaluate and optimize PCCT for cardiac imaging in patients during left ventricular assistance device (LVAD) therapy by conducting an in‐depth objective assessment of metal artifacts and visual grading.

Methods

Various scan and reconstruction settings were tested on a phantom and further evaluated on a patient acquisition to identify the optimal protocol settings. The phantom comprised an empty thoracic cavity, supplemented with heart and lungs from a cadaveric lamb. The heart was implanted with an LVAD (HeartMate 3) and iodine contrast. Scans were performed on a PCCT (NAEOTOM Alpha, Siemens Healthcare). Metal artifacts were assessed by three objective methods: Hounsfield units (HU)/SD measurements (Diff HU and SD ARTIFACT ), Fourier analysis (AmplitudeLowFreq), and depicted LVAD volume in the images (BloomVol). Radiologists graded metal artifacts and the diagnostic interpretability in the LVAD lumen, cardiac tissue, lung tissue, and spinal cord using a 5‐point rating scale. Regression and correlation analysis were conducted to determine the assessment method most closely associated with acquisition and reconstruction parameters, as well as the objective method demonstrating the highest correlation with visual grading.

Results

Due to blooming artifacts, the LVAD volume fluctuated between 27.0 and 92.7 cm 3 . This variance was primarily influenced by kVp, kernel, keV, and iMAR ( R 2 = 0.989). Radiologists favored pacemaker iMAR, 3 mm slice thickness, and T3D keV and kernel Bv56f for minimal metal artifacts in cardiac tissue assessment, and 110 keV and Qr40f for lung tissue interpretation. The model adequacy for Diff HU SD ARTIFACT , AmplitueLowFreq, and BloomVol was 0.28, 0.76, 0.29, and 0.99 respectively for phantom data, and 0.95, 0.98, 1.00, and 0.99 for in‐vivo data. For in‐vivo data, the correlation between visual grading (VG SUM ) and Diff HU SD ARTIFACT , AmplitueLowFreq, and BloomVol was −0.16, −0.01, −0.48, and −0.40 respectively.

Conclusion

We found that optimal scan settings for LVAD imaging involved using 120 kVp and IQ level 80. Employing T3D with pacemaker iMAR, the sharpest allowed vascular kernel (Bv56f), and VMI at 110 keV with kernel Qr40 yields images suitable for cardiac imaging during LVAD‐therapy. Volumetric measurements of the LVAD for determination of the extent of blooming artifacts was shown to be the best objective method to assess metal artifacts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsHeart-Assist DevicesPhantoms, ImagingTomography, X-Ray ComputedMetalsPhotonsHumansImage Processing, Computer-AssistedSheepAnimalsHeartRadiographic Image Interpretation, Computer-Assisted

Stichwörter der Autor:innen

heartleft ventricular assistance devicemetal artifactmyocardiumphoton‐counting computed tomography

Substanzen

Metals

Autor:innen und Institutionen

  • Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.
  • Center for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden.
  • Department of Radiology, Vestfold Hospital, Tønsberg, Norway.
  • Department of Thoracic and Vascular Surgery in Östergötland, and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.
  • Department of Radiology in Linköping, and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.

Förderung

  • CMIV
  • Swedish Research Council
  • Swedish Heart and Lung Foundation
  • ALF Grants Region Östergötland
  • Sweden's Innovation Agency VINNOVA
  • Foundation Linköpings Läkaresällskap Research Fund
  • Vetenskapsrådet

Zitieren

APA

Konst, B., Ohlsson, L., Henriksson, L., Sandstedt, M., Persson, A., & Ebbers, T. (2024). Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in‐depth assessment of metal artifacts. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14386

Vancouver

Konst B, Ohlsson L, Henriksson L, Sandstedt M, Persson A, Ebbers T. Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in‐depth assessment of metal artifacts. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14386

BibTeX
@article{Konst2024,
  author = {Konst, Bente and Ohlsson, Linus and Henriksson, Lilian and Sandstedt, Mårten and Persson, Anders and Ebbers, Tino},
  title = {Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in‐depth assessment of metal artifacts},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14386}
}
Kostenlos starten →