EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

High‐speed bioimpedance‐based gating system for radiotherapy: Prototype and proof of principle

Mikhail A Belikhin, Alexander P Chernyaev, Alexander A Pryanichnikov iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14491 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Mikhail A Belikhin und 2 weiteren Autor:innen. Beteiligte Einrichtung: JSC Protom, Protvino, Russian Federation. Hauptthemen laut MEDLINE: Electric Impedance, Radiotherapy Planning, Computer-Assisted, Signal-To-Noise Ratio. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate a novel bioimpedance‐based respiratory gating system (BRGS) designed for external beam radiotherapy and to evaluate its technical characteristics in comparison with existing similar systems. Materials and methods The BRGS was tested on three healthy volunteers in free breathing and breath‐hold patterns under laboratory conditions. Its parameters, including the time delay (TD) between the actual impedance change and the gating signal, temperature drift, root mean square (RMS) noise, and signal‐to‐noise ratio (SNR), were measured and analyzed.

Results

The gate‐on TD and the gate‐off TD were found to be 9.0 ± 2.0 ms [mean ± standard deviation (M ± SD)] and 7.2 ± 1.3 ms, respectively. The temperature drift of the BRGS output signal was 0.02 Ω after 30 min of operation. RMS noise averaged 0.14 ± 0.05 Ω (M ± SD) for all subjects and varied from 0.08 to 0.20 Ω with repeated measurements. A significant difference in SNR ( p < 0.001) was observed between subjects, ranging from 4 to 15.

Conclusion

The evaluated bioimpedance‐based gating system showed a high performance in real‐time respiratory monitoring and may potentially be used as an external surrogate guidance for respiratory‐gated external beam radiotherapy. Direct comparison with commercially available systems, 4D correlation studies, and expansion of the patient sample are goals for future preclinical studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Electric ImpedanceRadiotherapy Planning, Computer-AssistedSignal-To-Noise RatioHumansRadiotherapy DosageHealthy VolunteersRespirationMaleNeoplasmsFemaleAdultRespiratory-Gated Imaging TechniquesBreath HoldingRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

respiratory gated radiotherapythoracic bioimpedancetumor motion

Autor:innen und Institutionen

  • JSC Protom, Protvino, Russian Federation.
  • Lomonosov Moscow State University, Moscow, Russian Federation.
  • Division of Biomedical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.

ORCID-Profile: Alexander A Pryanichnikov

Zitieren

APA

Belikhin, M. A., Chernyaev, A. P., & Pryanichnikov, A. A. (2024). High‐speed bioimpedance‐based gating system for radiotherapy: Prototype and proof of principle. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14491

Vancouver

Belikhin MA, Chernyaev AP, Pryanichnikov AA. High‐speed bioimpedance‐based gating system for radiotherapy: Prototype and proof of principle. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14491

BibTeX
@article{Belikhin2024,
  author = {Belikhin, Mikhail A and Chernyaev, Alexander P and Pryanichnikov, Alexander A},
  title = {High‐speed bioimpedance‐based gating system for radiotherapy: Prototype and proof of principle},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14491}
}
Infokarte | EUVIMED